Rozonik HC2A-S humidity probeA probe, transmitter, or recorder used to measure the air environment. HC2A-S is a fully functional probe from Rotronic and the foundation of our product line. It measures humidity and temperature, calculates dew point and frost point.
Application:
HVAC, food, building equipment, paper, textile, pharmaceutical and other industries
Features:
Measure relative humidity and temperature, calculate dew point
• Accurate precision and repeatability
Excellent long-term stability<1% RH/year
• Integrated data collection and calibration recording functions
• Automatic diagnosis and temperature compensation
• Programmable alarm function
Advanced and easy-to-use calibration function
100% on-site exchange
Rozonik HC2A-S humidity probeTechnical Specifications
| Precision, high-precision mode | ± 0.5% RH/± 0.1 ° C, at 10... 30 ° C |
| At 23 ° C and 10, 20, 30, 40, 50, 60, 70, 80, 90% RH | |
| Precision, standard precision mode | ± 0.8% RH/± 0.1 ° C, at 10... 30 ° C |
| Calibration point at 23 ° C and 10, 35, 80% RH | |
| Probe type | Standard probe |
| scope of work | -50…100 °C / 0…100 %rh |
| Power supply/power consumption | 3.3…5 VDC |
| long-term stability | <1% RH/year |
| sensor | Hygromer HT-1 |
| temperature sensor | PT100 1/3 DIN Class B |
| Probe protection | Polycarbonate plastic cover, black |
| filter element | polyethylene |
| Filter type | Polyethylene, 40 μ m |
| Response time T 63 | <15 sec. |
| High wind speed | 20 m/s, With filter |
| computational load | Dew point or frost point |
| current consumption | 4.5 mA, 3.3 VDC power supply |
| Repeatability of AIRCHIP | 0.02 %rh / 0.01 °C |
| develop software | ROTRONIC HW4 |
| Adjust the probe through software | 1 point&multiple points% rh&° C |
| SOFTWARE UPDATE | Through the Internet and HW4 software |
| User information | From HW4 (Device Management) |
| Data processing through HW4 | Interface cable |
| The device is password protected | yes |
| User defined analog output | have |
| Analog output signal (standard) | 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C |
| Digital output signal (standard) | UART |
| Output signal type | 2 x 0…1 VDC |
| Shell material | polycarbonate |
| housing material | polycarbonate |
| cable length | Passive cables can be up to 5m long, while active cables can be up to 100m long |
| Probe material | PEEK |
| Reverse voltage protection | Mechanical connection |
| Service interface | UART |
| FDA/GAMP compatibility | Compatible with FDA 21 CFR Part 11 and GAMP |
| size | Ø15 x 108 mm |
| weight | 17 g |