SLT-08A-B Nuclear Medicine Decay Pool Control System

summary:
The radiation monitoring module adoptsThe SLT-08A-B instrument is a new type of online continuous measurement and alarm device for x/γ radiation. The host uses a 32-bit high-speed embedded microprocessor and a 1024x600 color LCD touch screen. The probe is designed with an IP68 waterproof structure and can be used for radiation monitoring in decay tanks. The host and probe adopt a reliable RS485 communication method, ensuring that the host can display the specific dose rate of each probe in real time and indicate the current status of the probe (online/offline/alarm) in real time. The display interface is beautiful and easy to operate.
The module is mainly used forThe measurement and alarm of dose rate using X or gamma rays are widely used in radioactive waste storage, industrial non-destructive testing, hospital gamma knife treatment, isotope application, gamma irradiation, hospital X-ray diagnosis, cobalt treatment, nuclear power plants, decay pools and other radioactive sites.
Functional Features
Øselect32-bit high-speed embedded microprocessor ensures real-time response to monitoring data and status.
Øuse7.0-inch 1024 × 600 color LCD touch screen, full touch operation, beautiful interface, strong operability.
ØThis module can be mounted simultaneouslyTwo waterproof probes can also automatically recognize and read the corresponding parameters on the host.
ØThe alarm threshold is continuously adjustable (the factory default setting for high threshold is5.0μSv/h, Low threshold of 0.0 μ Sv/h.
ØReal time display of probe status: online/Offline/alarm.
The host can be set with multiple alarm modes, supporting four modes: sound and light alarm, sound alarm, alarm light, and shutdown.
The host can automatically record threshold alarms and store them automatically, without losing them after power failure.
Ø The threshold alarm records of each probe can be manually queried.
ØCan be set/Read the drainage cycle of the decay pool and display the current decay time in real time;
ØDisplayableThe drainage status of Pool 2 (decay period) and Pool 3 (radiation exceedance value);
ØIt has real-time clock function, which can display year, month, day, minute, and second, and can also record real-time clock in case of power failure.
ØCommunication method: StandardRS485 interface, with a transmission distance of up to 800 meters.
Ø Customizable development according to needs
TECHNICAL INDEX
1. Detector: GM detector (energy compensation).
2. Measurement range: 0.01 μ Sv/h to 1Sv/h.
3. Sensitivity: 1 μ Sv/h>5CPS.
4. Response time: 1S (processed by algorithm).
5. Relative inherent error: ≤± 15%.
6. Energy response: 48Kev~3Mev ≤± 30%.
7. Power supply: 220V mains power.
8. Operating environment: Temperature range of -10 ℃ to+50 ℃, relative humidity (at 40 ℃) ≤ 90%.
9. Waterproof level of probe: IP68.
10. Size, host: 260 × 160 × 68mm (excluding base); Waterproof probe: φ 140 × 90 mm.
Water pump control module
The water pump control module adoptsThree or more intelligent controller hosts for water pumps, each controlling the water pump switches of three or more decay tanks. The intelligent controller of the water pump is based on microcomputer microcontroller technology, which integrates motor comprehensive protection and digital display functions. Its unique current and water level induction technology can achieve dual protection for water pump dry running and locked running due to water shortage.
The intelligent controller of the water pump can be connected to five types of control components: pressure transmitter, remote pressure gauge, electrical contact, float, and probe. In the application of decay pools,Considering the use of float level switches for sediment control in the first decay tank, dry reed level switches are used in the second and third decay tanks to achieve automated and intelligent control of the decay tanks.
Functional Features
1. Power on: Connect the power and select the desired control mode - water level control or pressure control.
2. Set current protection parameters: After starting the water pump, the current screen will display the actual working current of the water pump; Set the overload protection and no-load protection to the appropriate protection range, and the current range can also be automatically set through the one click setting function (manual setting is recommended).
3. Choose between automatic and manual working modes (switch with buttons):
4. Fault alarm: After the water pump is protected and stopped, an alarm will be issued, and the corresponding fault indicator light will be on;
5. Overvoltage protection: When the power supply voltage exceeds or falls below the voltage protection range, the power will be automatically cut off within 3 seconds.
6. Overload protection: The more the set value is exceeded, the faster the overload protection action will be.
7. No load protection: It has the characteristic of inverse time limit protection, that is, the more it is lower than the set value, the faster the no-load protection action will be.
Parameters
1. Static power consumption of the host:<3W.
2. Power supply voltage: AC380V ± 10% 50HZ;
3. Three levels of water level display: low, medium, high, in sequence
4. Maximum control distance of water level: ≤ 500 meters in environments without strong electric and magnetic fields or interference;
5. Overvoltage protection range: 260V (≤ 3-second trip)