Overview

Product application: Differential Scanning Calorimetry (DSC) measures the relationship between temperature and heat flux related to thermal transformation inside materials. It has a wide range of applications, especially in material research and development, performance testing, and quality control

Product Details

Product Application
Differential scanning calorimeterDSCIt measures the relationship between temperature and heat flux related to thermal transformation inside materials, and has a wide range of applications, especially in material research and development, performance testing, and quality control. The characteristics of materials, such as glass transition temperature, cold crystallization, phase transition, melting, crystallization, product stability, and solidification/Crosslinking, oxidation induction period, etc.

Product Features

  • Tzerotechnology:Tzero™ Sensor, eliminate the originalDSCThe thermal effects that affect baseline stability, resolution, and sensitivity in the design result in more reliable heat flux signals. The uniquely designed reference temperature sensor and measurement circuit can detect and compensate for the impedance and heat capacity imbalances caused by these thermal effects.
  • Advanced modulationMDSCTechnology, direct measurement of heat capacityCp
  • Touch screen, replaceableDSCStove, digital flow control50Automatic sampler, automatic loading furnace cover


Technical Specifications

  • Temperature range:-180725
  • Temperature accuracy: ±0.01
  • Temperature accuracy: ±0.01
  • Dynamic measurement range:>±500mW
  • Calorimetry accuracy (indium standard): ±0.05%
  • Calorimetric repeatability (indium standard): ±0.05%
  • Sensitivity:0.2MW
  • TzeroBaseline curvature and repeatability:10MW
  • Indium peak height/full width at half maximum:60mW/
  • Relative resolution:2.9
  • PLATIUMsoftware