E-mail
admin@xian-victor.com
- Phone
-
Address
25th Floor, Building G, Wangzuo Modern City, No. 35 Tangyan Road, High tech Zone, Xi'an City
Xi'an Shengli Instrument Co., Ltd
admin@xian-victor.com
25th Floor, Building G, Wangzuo Modern City, No. 35 Tangyan Road, High tech Zone, Xi'an City
Temperature calibrator VICTOR 07It is a portable device designed for on-site circuit debugging. It can provide 24V circuit power supply and measure circuit current, making it easy to debug on-site transmitters; The active/passive current output function facilitates the debugging of systems such as PLC and DCS.
Temperature calibrator VICTOR 07It has been widely applied in fields such as petrochemicals, biopharmaceuticals, industrial automation, colleges and research institutions.












| Input function output function |
DCV | DCI | OHM |
|
TC | RTD |
| DCV | ● | ● | ● | ● | ● | ● |
| DCI | ● | ● | ● | ● | ● | ● |
| OHM | ● | ● | x | ● | ● | x |
| mA TEMP | ● | ● | ● | x | x | x |
| TC | ● | ● | ● | x | x | x |
| RTD | ● | ● | x | x | x | x |
| output function | range | output range | resolution | accuracy | Remarks | |
| DC voltage DCV |
100mV | -10.00mV~110.00mV | 0.01mV | 0.05%+0.03mV | Maximum output current: 0.5mA | |
| 1V | -0.1000V~1.1000V | 0.0001V | 0.05%+0.3mV | Maximum output current 2mA | ||
| 10V | -1.000V~11.000V | 0.001V | 0.05%+2mV | Maximum output current 5mA | ||
| direct current DCI |
20mA | 0.000mA~22.000mA | 0.001mA | 0.05%+4uA | At 20 mA, the maximum load resistance is 1000 Ω When simulating a transmitter, the external circuit provides a power supply of 5-28V |
|
| ohm OHM |
400Ω | 0~400.0Ω | 0.1Ω | 0.05%+0.2Ω | The excitation current is ± 0.1 to ± 5mA Maximum output voltage 2V When the excitation current is ± 0.1-0.5mA, add an additional error of 0.1 Ω Accuracy does not include lead resistance |
|
| 4KΩ | 0~4.000KΩ | 1Ω | 0.05%+2Ω | |||
| thermocouple TC |
R | 0°C~1767°C | 1°C | 0~100℃ 10~1767℃ 0~100℃ 100~1767℃ |
1.5℃ 1.2℃ 1.5℃ 1.2℃ |
Using ITS-90 temperature scale Accuracy does not include errors in cold end compensation Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
| S | 0°C~1767°C | |||||
| K | -200.0°C~1372.0°C | 0.1°C | -200.0~-100.0℃ -100.0~400.0℃ 400.0~1200.0℃ 1200.0~1372.0℃ |
0.6℃ 0.5℃ 0.7℃ 0.9℃ |
||
| E | -200.0°C~1000.0°C | -200.0~-100.0℃ -100.0~600.0℃ 600.0~1000.0℃ |
0.6℃ 0.5℃ 0.4℃ |
|||
| J | -200.0°C~1200.0°C | -200.0~-100.0℃ -100.0~800.0℃ 800.0~1200.0℃ |
0.6℃ 0.5℃ 0.7℃ |
|||
| T | -250.0°C~400.0°C | -250.0~400.0℃ | 0.6℃ | |||
| N | -200.0°C~1300.0°C | -200.0~-100.0℃ -100.0~900.0℃ 900.0~1300.0℃ |
1.0℃ 0.7℃ 0.8℃ |
|||
| B | 600°C~1820°C | 1°C | 600~800℃ 800~1820℃ |
1.5℃ 1.1℃ |
||
| A | 0°C~2500°C | 1°C | ||||
| C | 0°C~2310°C | 1°C | ||||
| D | 0°C~2310°C | 1°C | ||||
| thermal resistor RTD |
Pt100 385 |
-200.0°C~800.0°C | 0.1°C | -200.0~0.0℃ 0.0~400.0℃ 400.0~800.0℃ |
0.3℃ 0.5℃ 0.8℃ |
Using Pt (385) temperature scale The excitation current is ± 0.1 to ± 5mA Maximum output voltage 2V When the excitation current is ± 0.1-0.5mA, add an additional error of 0.1 Ω Accuracy does not include lead resistance Accuracy does not include the influence of thermoelectric potential |
| Pt200 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.8℃ 0.9℃ 1.0℃ |
|||
| Pt500 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.4℃ 0.5℃ 0.7℃ |
|||
| Pt1000 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.2℃ 0.5℃ 0.7℃ |
|||
| Cu50 | -50.0°C~150.0°C | 0.6℃ | ||||
| Measurement function | range | measurement range | resolution | accuracy | Remarks | |
| DC voltage DCV |
100mV | -110.000mV~110.000mV | 0.001mV | 0.05%+0.03mV | Input resistance: approximately 1M Ω | |
| 1V | -1.1000V~1.1000V | 0.0001V | 0.05%+0.3mV | |||
| 35V | -35.000V~35.000V | 0.001V | 0.02%+2mV | |||
| direct current DCI |
30mA | -35.000mA~35.000mA | 0.001mA | 0.02%+4uA | Shunt resistance: approximately 10 Ω Input resistance: approximately 20 Ω |
|
| resistance OHM |
500Ω | 0~550.00Ω | 0.01Ω | 0.05%+0.2Ω | 2W/3W/4W measurement 500 Ω approximately 1mA excitation 5K Ω approximately 0.1mA excitation Open circuit voltage: approximately 2.5V The measurement accuracy does not include lead resistance |
|
| 5KΩ | 0~5.5000KΩ | 0.0001KΩ | 0.05%+2Ω | |||
| thermocouple TC |
R | 0°C~1767°C | 1°C | 0~500℃ 500~1767℃ |
2.5℃ 2℃ |
Using ITS-90 temperature scale Accuracy does not include errors in cold end compensation Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
| S | 0°C~1767°C | |||||
| K | -100.0°~1372.0°C | 0.1°C | -100.0~0.0℃ 0.0~1372.0℃ |
1.2℃ 0.8℃ |
||
| E | -50.0°C~1000.0°C | -50.0℃~0.0℃ 0.0~1000.0℃ |
0.9℃ 1.5℃ |
|||
| J | -60.0°C~1200.0°C | -60.0~0.0℃ 0.0~1200.0℃ |
1.0℃ 0.7℃ |
|||
| T | -100.0°C~400.0°C | -100.0~0.0℃ 0.0~400.0℃ |
1.0℃ 0.7℃ |
|||
| N | -200.0°~1300.0°C | -200.0~0.0℃ 0.0~1300.0℃ |
1.5℃ 0.9℃ |
|||
| B | 600°C~1820°C | 1°C | 600~800℃ 800~1000℃ 1000~1820℃ |
2.2℃ 1.8℃ 1.4℃ |
||
| A | 0°C~2500°C | 1°C | ||||
| C | 0°C~2310°C | 1°C | ||||
| D | 0°C~2310°C | 1°C | ||||
| thermal resistor RTD (4W) |
Pt100 385 |
-200.0°C~800.0°C | 0.1°C | -200.0~0.0℃ 0.0~400.0℃ 400.0~800.0℃ |
0.5℃ 0.7℃ 0.8℃ |
Using Pt (385) temperature scale 2W/3W/4W measurement 500 Ω approximately 1mA excitation 5K Ω approximately 0.1mA excitation Open circuit voltage: approximately 2.5V The accuracy does not include the error caused by the mismatch of 2 W/3 W measurement lead resistance Accuracy does not include the inaccuracy of sensors Accuracy does not include the influence of thermoelectric potential |
| Pt200 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.8℃ 0.9℃ 1.0℃ |
|||
| Pt500 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.4℃ 0.5℃ 0.7℃ |
|||
| Pt1000 385 |
-200.0°C~630.0°C | -200.0~100.0℃ 100.0~300.0℃ 300.0~630.0℃ |
0.3℃ 0.5℃ 0.7℃ |
|||
| Cu50 | -50.0°C~150.0°C | -50.0~150.0℃ | 0.7℃ | |||
| On/off detection | 500Ω | ≤ 50 Ω sound output | 0.01Ω | About 1mA incentive | ||
| loop-powered LOOP |
24 V | 10% | short circuit protection Maximum current: 22 mA Maximum input voltage: 60 V DC |
|||
3、 General characteristics of the product
*Working temperature and humidity: 0 to 50 º C ≤ 80% RH, no condensation; 40 to 50 º C ≤ 70% RH.
*Storage temperature and humidity: -25 to 60 º C ≤ 90% RH, no condensation.
*Electrical safety: EN61010-1:2001
Voltage endurance: AC3540V (50/60Hz)/5 seconds between the terminal and the casing.
Insulation impedance: DC1000V/100M Ω or greater between the terminal and the casing.
*Electromagnetic compatibility (EMC): EN61326-1:2006
Meets performance criterion 2, which means that the functionality and performance are temporarily reduced or lost, but can be restored on their own.
*Protection level IP40.
*Vibration and Drop: IEC 60068-2-64:2008, IEC 60068-2-32:2008
Random, 2g, 5-500Hz; 1-meter drop test.
*CE certification.
*Quality standards: Developed, designed, and manufactured in accordance with TUV ISO 9001.
*Calibration cycle and preheating time:
To ensure the accuracy of this table, its calibration cycle is one year
The startup preheating time is greater than 30 minutes
*Instrument display and buttons
Segment code LCD display, white backlight
Adopting silicone rubber buttons
*Instrument power supply
Powered by 3 5th alkaline batteries