Overview of Total Copper Online Automatic Analyzer (PH-W02-TCu01): The total copper online automatic analyzer is an optical heavy metal online analyzer designed in accordance with national environmental protection standards. It is used for simultaneous online analysis of multiple heavy metal elements in water quality such as surface water, industrial process water, and industrial wastewater. The measured heavy metal elements include Class I and Class II heavy metal prevention and control projects in the Ministry of Environmental Protection's "Twelfth Five Year Plan" for comprehensive prevention and control of heavy metal pollution
Total Copper Online Automatic Analyzer (PH-W02-TCu01)

Overview:
The total copper online automatic analyzer is an optical method heavy metal online analyzer designed in accordance with national environmental protection standards. It is used for simultaneous online analysis of multiple heavy metal elements in water quality such as surface water, industrial process water, and industrial wastewater. The measured heavy metal elements include Class I and Class II heavy metal prevention and control projects in the Ministry of Environmental Protection's "Twelfth Five Year Plan" for comprehensive prevention and control of heavy metal pollution.
Product Principle:
This study used the sodium diethyldithiocarbamate spectrophotometric method to determine total copper. At high temperatures, various copper compounds in the water sample are oxidized and digested by nitric acid to form divalent copper ions. Under ammonia conditions, copper ions react with sodium diethyldithiocarbamate to form yellow brown complexes. Within the concentration range, the absorbance of the system is proportional to the concentration of copper ions, and the total copper content in the water sample can be determined by spectrophotometry.
Application scenarios:
Suitable for online monitoring of surface water in textile printing and dyeing, petrochemical, coking, meat processing, fertilizer industry wastewater, rivers, lakes, etc.
Features:
1. Stable performance of monochromatic light element with good repeatability in measurement;
2. Multiple modes such as scheduled collection, equidistant collection, and remote control collection can be scheduled;
3. Visual system action flow or measurement curve analysis process;
4. Automatic sample retention interface for excessive data, providing convenience for future reference and retesting;
5. Innovative data bus distributed module reduces system maintenance rate;
6. Automatic reset and maintenance require minimal human intervention;
7. Network devices have remote access and automatic fault diagnosis functions;
8. Master slave data port, which can be configured with network or any wireless module;
9. 10inchTFTTouch screen, harmonious operation and data display interface;
10. spacingUSBThe data export function allows for the selection of data export time periods;
11. Friendly graphical and tabular human-machine interface, easy to operate equipment.
Technical Specifications:
Scope of Application |
This product is suitable for(0~10)mg/LWater quality within the scope |
test method |
Sodium diethyldithiocarbamate spectrophotometry |
measurement range |
0-3.00mg/LThe0.10-10.00mg/L |
resolution |
0.001 |
accuracy |
≥0.2 mg/LWhen, not exceeding ±10%;<0.2 mg/LWhen, not exceeding ±0.02 mg/L; |
repetitiveness |
≤5% |
zero drift |
±0.5mg/L |
range drift |
24hInternally ≤±10% |
measurement period |
The minimum measurement period is30Minutes, according to the on-site environment, can be5~120minArbitrarily modify the color rendering time; |
sampling period |
Time interval (adjustable), hourly or trigger measurement mode, adjustable |
calibration cycle |
Automatic calibration(1~99Adjustable according to the actual water sample, manual calibration can be set |
maintenance cycle |
The maintenance interval is greater than1Month, every appointment30min |
reagent consumption |
Generally less than0.5yuan/sample |
Human machine operation |
Touch screen display and command input |
Self checking protection |
Self diagnosis of instrument working status, no data loss after abnormalities or power outages, automatic elimination of residual reactants and automatic restoration of operation after abnormal reset or power outages and incoming calls |
data storage |
not less than1Annual data storage |
input interface |
discrete signal |
output interface |
2roadRS232Digital output,2road4-20mAanalog output |
working environment |
Indoor work, recommended temperature15-25℃, humidity ≤90%(No condensation) |
Power supply and power consumption |
AC230±10%V,50~60Hz,5A |
size |
tall1500*wide500*deep450(mm) |