Overview of Total Manganese Online Automatic Analyzer (PH-W02-TCu01): The total manganese online automatic analyzer is an optical heavy metal online analyzer designed in accordance with national environmental protection standards. It is used for simultaneous online analysis of multiple heavy metal elements in water quality such as surface water, industrial process water, and industrial wastewater. The measured heavy metal elements include Class I and Class II heavy metal prevention and control projects in the Ministry of Environmental Protection's "Twelfth Five Year Plan" for comprehensive prevention and control of heavy metal pollution
Total Manganese Online Automatic Analyzer (PH-W02-TCu01)

overview:
The total manganese online automatic analyzer is an optical method heavy metal online analyzer designed in accordance with national environmental protection standards. It is used for simultaneous online analysis of multiple heavy metal elements in water quality such as surface water, industrial process water, and industrial wastewater. The measured heavy metal elements include Class I and Class II heavy metal prevention and control projects in the Ministry of Environmental Protection's "Twelfth Five Year Plan" for comprehensive prevention and control of heavy metal pollution.
Product principle:
The water sample is dissolved at high temperature by an oxidant, and after cooling, it forms a colored complex with a color developing agent. The analyzer detects the change in color and converts this change into the total manganese value for output. The amount of colored complex generated is equivalent to the total manganese content.
Application scenarios:
Suitable for online monitoring of surface water in textile printing and dyeing, petrochemical, coking, meat processing, fertilizer industry wastewater, rivers, lakes, etc.
characteristic:
1. Stable performance of monochromatic light element with good repeatability in measurement;
2. Multiple modes such as scheduled collection, equidistant collection, and remote control collection can be scheduled;
3. Visual system action flow or measurement curve analysis process;
4. Automatic sample retention interface for excessive data, providing convenience for future reference and retesting;
5. Innovative data bus distributed module reduces system maintenance rate;
6. Automatic reset and maintenance require minimal human intervention;
7. Network devices have remote access and automatic fault diagnosis functions;
8. Master slave data port, which can be configured with network or any wireless module;
9. 10cunTFTTouch screen, harmonious operation and data display interface;
10. spacingUSBThe data export function allows for the selection of data export time periods;
11. Friendly graphical and tabular human-machine interface, easy to operate equipment.
Technical parameters:
scope of application |
This product is suitable for(0~5)mg/LWater quality within the scope |
test method |
Potassium periodate spectrophotometric method |
Measurement range |
0.00-5.00 mg/L |
resolution |
0.001 |
accuracy |
≥0.2 mg/LWhen, not exceeding ±10%;<0.2 mg/LWhen, not exceeding ±0.02 mg/L; |
repeatability |
≤5% |
zero drift |
±0.5mg/L |
Span Drift |
24hInternally ≤±10% |
Measurement period |
The minimum measurement period is30Minutes, according to the on-site environment, can be5~120minArbitrarily modify the color rendering time; |
sampling period |
Time interval (adjustable), hourly or trigger measurement mode, adjustable |
calibration cycle |
Automatic calibration(1~99Adjustable according to the actual water sample, manual calibration can be set |
Maintenance Cycle |
The maintenance interval is greater than1Month, every appointment30min |
Reagent consumption |
Generally less than0.5first/sample |
human-machine operation |
Touch screen display and command input |
Self checking protection |
Self diagnosis of instrument working status, no data loss after abnormalities or power outages, automatic elimination of residual reactants and automatic restoration of operation after abnormal reset or power outages and incoming calls |
Storage |
not less than1Annual data storage |
Input Interface |
Switching quantity |
output interface |
2roadRS232Digital output,2road4-20mAAnalog Output |
work environment |
Indoor work, recommended temperature15-25℃, humidity ≤90%(No condensation) |
Power supply and power consumption |
AC230±10%V,50~60Hz,5A |
size |
tall1500*wide500*deep450(mm) |