Volatile Organic Gas Online Continuous Monitoring System/Overview In response to the serious environmental air pollution in China and the urgent need to monitor volatile organic gas emissions from fixed pollution sources such as petrochemicals, electronics, and industrial coatings, our company has launched the independently developed HV-3060 volatile organic gas online continuous monitoring system

Product Details

Volatile Organic Gas Online Continuous Monitoring System/Overview

In response to the severe environmental air pollution in China and the urgent need to monitor volatile organic gas emissions from fixed pollution sources in key polluting industries such as petrochemicals, electronics, and industrial coatings, our company has launched the independently developed HV-3060 volatile organic gas online continuous monitoring system. The volatile gas online continuous monitoring system is an online monitoring system based on gas chromatography technology, which can monitor volatile organic compounds (including methane, non methane total hydrocarbons, and benzene derivatives) in flue gas online, and can also expand other components (such as ketones, lipids, and other organic compounds) according to user needs. It has high measurement accuracy and a wide dynamic range. The continuous online monitoring system for volatile organic gases adopts intelligent high-performance gas sensors, which have the characteristics of easy installation, high measurement accuracy, good reliability, fast response speed, and strong anti-interference ability

Volatile Organic Gas Online Continuous Monitoring System/principle

挥发性有机物在线连续监测系统

The sampling system of the volatile organic gas continuous monitoring instrument extracts the measured gas from the sampling point, and after dust removal by the high-temperature sampling probe, it enters the online gas chromatograph through the high-temperature tracing pipeline. The chromatograph is equipped with a built-in heating box, which keeps all the pipelines through which the sample passes at high temperature. Then, a high-sensitivity hydrogen flame ionization detector (FID) is used to detect the sample, and finally, data collection, analysis, processing, transmission, and storage are automatically completed through workstation software.

Volatile Organic Gas Online Continuous Monitoring System/feature

  • High temperature control accuracy

    Industrial rapid chromatography design, high temperature control accuracy, high reliability, long service life, can cope with various harsh and complex working conditions, fully meeting the requirements of HJ1013-2018 new standards.

  • Original imported high-temperature VICI injection valve

    The volatile organic gas online continuous monitoring instrument adopts the original imported high-temperature VICI injection valve, with no dead volume injection and no sample adsorption.

  • Ppb level

    The hydrogen flame ionization detector for trace components has a detection limit of up to ppb level.

  • Customized sampling system

    The sampling pipeline of the volatile organic gas online monitoring instrument is designed with high temperature throughout the process, without any adsorbent material, ensuring the purity of the sample. We can customize sampling systems according to customer requirements, including positive pressure design, negative pressure design, and dilution design.

  • Simple operation and maintenance

    The online continuous monitoring instrument for volatile organic gases adopts a configuration design, which is easy to operate and maintain. It is easy for operation and maintenance personnel to learn and understand, and does not require professional chromatographic knowledge.

  • Imported components

    The volatile organic gas online monitoring and pretreatment system adopts imported components, solenoid valves, sampling pumps, and has a low repair rate, ensuring the stability of the equipment.

  • Customization of monitoring factors

    The volatile organic gas online monitoring instrument can accept customized monitoring factors from customers, support dual channel dual analyzer configuration, and directly calculate the removal rate of treatment equipment.

  • Equipped with a six channel full process heat tracing system

    The volatile organic gas online monitoring device can be paired with a six channel full process heat tracing system to simultaneously detect the concentration of six channels.

Volatile Organic Gas Online Continuous Monitoring System/parameter

project

Parameters of Continuous Online Monitoring System for Volatile Organic Compounds Gas

Detection factor

Total hydrocarbons, methane, non methane total hydrocarbons, benzene derivatives, alcohols, lipids, halogenated hydrocarbons, etc

range

Methane 0-10/100/1000/10000/100000mg/m3 (optional); Non methane total hydrocarbon methane 0-10/100/1000/10000/100000mg/m3

detector

GC-FID Hydrogen Flame Ion Detector

detection limit

Methane ≤ 0.1ppm, non methane ≤ 0.05ppm, benzene ≤ 0.03ppm

repetitiveness

RSD≤2%.

Analysis cycle

NMHC<2min Characteristics<10min

zero drift

≤±1%FS

range drift

≤±2%FS

Power supply

<500W,220V AC/50Hz

working environment

Temperature: (5-40) ℃, Humidity: (10% -90%) RH

Air source requirements

Carrier gas: high-purity nitrogen or zero grade air (≥ 99.999%); Combustion gas: high-purity hydrogen gas (≥ 99.999%) Auxiliary gas: zero grade air (hydrocarbons<20ppb)

output

4-20mA、RS232/RS485、 Ethernet

size

Height 1800 * Length 900 * Width 800 (mm)

Volatile Organic Gas Online Continuous Monitoring System/application

The HV-3060 volatile organic gas online continuous monitoring system can be widely used for monitoring toxic and harmful volatile gas emissions in various industries such as environmental protection, metallurgy, petroleum and petrochemical, and chemical engineering.