The Xilin injection bottle micro leakage sealing tester is specialized in non-destructive positive and negative pressure micro leakage testing of various empty/pre filled syringes, water and powder injection bottles (glass/plastic), filling and capping bottles, milk powder cans, other hard packaging containers, electrical components and other samples. This sealing instrument adopts advanced design and rigorous, scientific calculation methods to ensure its rapid testing, high accuracy, and high stability. It can also meet users' non-standard (software or testing fixture) customization needs.


Product Introduction:
Xilin Injection Bottle Micro Leakage Sealing TesterProfessional application for non-destructive positive and negative pressure micro leak testing of various empty/pre filled syringes, water and powder injection bottles (glass/plastic), filling and capping bottles, milk powder cans, other hard packaging containers, electrical components and other samples. This sealing instrument adopts advanced design and rigorous, scientific calculation methods to ensure its rapid testing, high accuracy, and high stability. It can also meet users' non-standard (software or testing fixture) customization needs.
| Comparison of Leakage Test Methods |
| Compare projects |
Water bath test |
Micro leak detection |
| The minimum detectable aperture |
10~20μm
|
<1 μ m detectable |
| Test time |
>30s |
About 10 seconds |
| Quantitative indication management |
Sensory testing, difficult to quantify
|
Leakage volume, leakage aperture |
| reproducibility |
Poor repeatability |
Can perform numerical reproduction |
| Destructive/Non destructive |
Destructive test |
Non destructive testing |
| test medium |
Water (requires maintenance) |
air |
| human error |
Small leaks are difficult to detect |
Numerical management, no need for manual labor |
| Leakage point confirmation |
Can locate the leakage point |
Cannot confirm the leakage point
|
Execution standard:
ASTM F2338-13 Standard Test Method for Packaging Leakage - Vacuum Attenuation Method USP1207 United States Pharmacopeia Standard
Drug GMP Guidelines - Sterile Drugs 11.1 Sealing Integrity Testing
Microbial examination method for Part Four of the 2020 edition of the Chinese Pharmacopoeia
Technical Guidelines for Sealing Research of Chemical Drug Injection Packaging Systems (Trial)
YYT 0681.18-2020 Test Methods for Aseptic Medical Device Packaging Part 18: Non destructive Testing of Packaging Leakage by Vacuum Attenuation Method
Chinese Pharmacopoeia 2025 Edition
GB/T 15171-2025 Test Method for Sealing Performance of Packaging Components
Technical advantages:
★ Optional built-in 10 inch touchscreen computer and external computer;
★ It can accurately display the leakage aperture (≥ 1 μ m) and leakage amount;
The test chamber and the host are arranged separately, and one set of test chambers is suitable for more than 5 specifications of samples;
★ The testing chamber is made of aluminum alloy or stainless steel, with pneumatic clamping;
★ Built in flow meter, one click completion of flow calibration;
★ Equipped with three calibration methods: zero point, leak hole, and flow rate;
The test results have three judgment modes: pressure attenuation, leakage aperture, and leakage flow rate
The test result shows that the flow error is less than or equal to 0.1 sccm
★ Vacuum resolution ≤ 1pa/0.01mbar/0.0001psi
★ Capable of converting testing units such as kpa/mbar/pa/psi
★ Can detect positive and negative pressure attenuation tests for various soft and hard samples such as penicillin bottles, infusion bags, contact lenses, milk powder cans, electronic accessories, etc;
Xilin Injection Bottle Micro Leakage Sealing Tester
Main parameters:
| Parameter Name/Equipment Model |
MLT -v100 |
MLT-V100(T) |
MLT-V100(P) |
| test method |
Vacuum attenuation method |
Vacuum attenuation method/pressure attenuation method |
| measurement and control technology |
Microcomputer, dual sensor testing technology |
| Vacuum method |
vacuum pump |
| Positive pressure method |
none |
External gas source |
| testing unit |
kPa/mbar/psi/Pa |
| resolution |
1Pa |
0.1Pa |
| test scope |
-99~0kPa |
-99~400kPa |
| measurement range |
≥0.01ccm (约1μm ) |
| test accuracy |
±0.2μm (≥3μm ) |
| test accuracy |
±0.3μm (<3μm ) |
| Master control chip |
STM microcomputer chip |
| sampling frequency |
50~200Hz |
| control method |
7 "LCD touch screen |
10.1 "LCD touch screen |
computer software |
| control system |
Microcontroller system |
Microcontroller system |
| audit trail |
audit trail |
| data storage |
600 sets |
no upper limit |
| data transmission |
USB/RS232/485 |
| keyboard and mouse |
not supported |
support |
| reporting method |
mini printer |
Micro printer/A4 report |
A4 Report |
| power supply |
220V 50Hz |
| air supply pressure |
none |
≥0.5MPa |
| air source interface |
6mm polyurethane pipe |
| Appearance material |
Engineering plastics/stainless steel |
| Overall dimensions |
320(L)×480(W)×320(H) |
340(L)×480(W)×320(H) |
340(L)×470(W)×320(H) |
| Equipment weight |
About 20kg |
| Fixture configuration |
Manual/automatic fixture |