1、 Performance indicators
1. Optical system
Wavelength range: 190nm-900nm, automatically adjustable
Grating line density: 1800 lines/mm
Monochromator: aberration reducing Czerny Turner type monochromator
Spectral bandwidth: 0.1, 0.2, 0.4, 1.0, 2.0nm, with multiple automatic switching modes
Wavelength accuracy: ± 0.10nm
Wavelength repeatability: 0.05nm
Baseline drift: static 0.002A/30min, dynamic 0.006A/30min
Detector: Wide spectral range photomultiplier tube
Resolution: Able to distinguish the dual lines of 279.5nm and 279.8nm in the manganese lamp, and the energy of the valley between the two spectral lines should be less than 20%
2. Light source system
Lamp holder: Automatic switching of 8 lamp holders
Lamp power supply method: 400Hz square wave pulse
Lamp current regulation range: 0-10mA average current
3. Atomization system
Detection limit (Cu): 0.003 μ g/ml
Burner: 100mm all titanium metal burner
Precision: RSD 0.5%
Spray: metal sheathed high-efficiency glass spray
Atomization chamber: corrosion-resistant all plastic atomization chamber
Safety measures: It has multiple automatic safety protection functions, such as acetylene leakage alarm and automatic shutdown system.
4. Flame analysis method
Measurement methods: standard curve method, standard addition method, interpolation method
Concentration calculation method: automatic fitting of working curve, automatic correction of sensitivity, automatic calculation of concentration and content.
Repeated measurement times: 1~99 times, automatically calculate the mean, content, standard deviation, and relative standard deviation.
Result printing: Print test data or final analysis report. The results can be exported as a Word worksheet.
Software environment: Windows XP/Windows 7 operating system, professional Chinese software
2、 Standard configuration
1. Host
2. Low noise oil-free air compressor
3. Hollow cathode lamp
4. Operating software
3、 Design Features of ZCA-1000AF Fully Automatic Flame Atomic Absorption Spectrophotometer
1. Total reflection optical system, full band chromatic aberration reduction, automatic wavelength scanning, and automatic peak searching.
2. The uniquely designed all titanium burner ensures smooth combustion and fast thermal equilibrium speed.
3. Automatic adjustment of negative high voltage, lamp current, optimal element lamp position, automatic ignition, and automatic shutdown protection. The automatic control system of the 8-lamp rack (optional with 2 high-performance lamps) can adjust the hollow cathode lamp to the optimal position in all directions, and can set the position of a specific element lamp separately. The work lamp and preheating lamp can be freely set to save the preparation time of the element lamp to the greatest extent possible.
4. It can be upgraded to a graphite furnace atomic absorption spectrophotometer.