Welcome Customer !

Membership

Help

Anhui Xun'er Instrument Technology Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Article

Anhui Xun'er Instrument Technology Co., Ltd

  • E-mail

    xrybkj@163.com

  • Phone

    15155078600

  • Address

    589 Tiankang Avenue, Tianchang City, Anhui Province

Contact Now
Calibration Specification for Armored Thermocouples
Date: 2025-07-07Read: 8
1、 Scope of application of the standard
Applicable object: Low cost metal armored thermocouples (K, N, E, J, T type, etc.) with metal casing length ≥ 500 mm.
Temperature measurement range: -40 ℃~1100 ℃.
Calibration object: Newly manufactured, in use, and repaired armored thermocouples. ‌
2、 Core metrological characteristic requirements
Room temperature insulation resistance
The insulation resistance between thermoelectric electrodes and between electrodes and metal sleeves is ≥ 1000 M Ω· m (20 ± 15 ℃, humidity ≤ 80%). ‌
Allowable deviation of indication
According to the division number and temperature range classification (1 level/2 levels), for example:
K-type (-40~375 ℃): ± 1.5 ℃ (Level 1) or ± 2.5 ℃ (Level 2)
E-type (-40~375 ℃): ± 1.5 ℃ (Level 1) or ± 2.5 ℃ (Level 2)
Please refer to Appendix E of the specification for the tolerance table
Thermal response time (dynamic performance)
Requirement: ≤ 180 seconds (conventional diameter), ultra-fine armor wire needs to be verified through a precision constant temperature bath. ‌
3、 Calibration conditions and equipment
Environmental requirements
Temperature: 15~35 ℃, fluctuation ≤ 1 ℃/hour; Humidity: ≤ 85% RH
Standard configuration
Main standard: second-class standard platinum resistance thermometer or standard platinum rhodium 10 platinum thermocouple.
Electrical measuring equipment: 0.02 level or above digital multimeter.
Temperature field source: High temperature furnace (temperature field uniformity ≤ 1 ℃/60mm) or constant temperature bath. ‌
Supporting tools
Freezing point thermostat (reference terminal 0 ℃ control), compensation wire (correction value needs to be calibrated). ‌
4、 Calibration method and process
Calibration point selection
Covering the process temperature range of the tested equipment, distribute points at intervals of 150 ℃ (such as 500 ℃, 650 ℃, 800 ℃).
Prohibition of extrapolation method: Deviation cannot be calculated outside the measured temperature range. ‌
Operation steps
① Preprocessing: Check the appearance and insulation resistance. New couples do not require annealing (experimental results have shown that annealing affects approximately 10% tolerance). ‌
② Layout: The standard and the calibrated couple are placed at the center of the temperature equalizing block/tube furnace, with an insertion depth of ≥ 300mm.
③ Measurement: After constant temperature (fluctuation ≤ 0.2 ℃/min), read the thermoelectric potential values in reverse order (4 times/point average). ‌
④ Deviation calculation:
5、 Result processing and resumption management
Calibration certificate content
Indication deviation, expanded uncertainty, environmental parameters (temperature and humidity/atmospheric pressure), traceability validity statement. ‌
Uncertainty assessment
Including standard error, temperature field uniformity, repeatability and other components (see Appendix G of the specification for examples)
Resumption period
Regular use: 6 months
High frequency/harsh environment: 3 months
After repairing or replacing core components: calibrate immediately. ‌