Prevention first "is the key to daily maintenance of cold and hot shock test chambers
Date: 2025-12-08Read: 4

thermal shock chamberAccurate operation relies on daily meticulous maintenance, among which defrosting, water tank cleaning, and basic maintenance are the core links. By doing these three things well, we can not only ensure temperature accuracy (± 0.5 ℃), but also extend equipment life by 3-5 years and reduce maintenance costs.

1、 Defrosting: Avoid icing that affects temperature efficiency

Freezing inside the low-temperature box can cause delayed temperature changes and increased energy consumption. It is necessary to trigger defrosting according to "ice thickness>5mm" or "every 150 hours of operation", and operate in both manual and automatic modes:

(1) Automatic defrosting (recommended): Set the "defrost cycle" in the control system”(default 24 hours) and "defrost duration" (10-15 minutes), the device will automatically heat up and defrost, and defrost water will flow into the water collection tray through the drainage pipe. The water collection tray needs to be checked and drained daily to avoid overflow.

(2) Manual defrosting (when icing is severe): First, stop the machine and take out the sample, then turn off the main power supply; When the temperature inside the box rises to 5-10 ℃, use the plastic defrosting shovel provided with the equipment to gently scrape the ice layer (metal tools are prohibited to prevent scratching the inner liner); After the ice melts, wipe the inner container with a dry cloth to ensure there are no residual water stains before restarting the equipment.

Attention: After defrosting, run the machine empty for 30 minutes. Wait for the temperature inside the box to stabilize before placing the sample to avoid temperature fluctuations that may affect the test data.

2、 Water tank cleaning: anti clogging, anti pollution

The humidification water tank (supply) and drainage water tank (storage) of the equipment are prone to scale and mold growth, and should be operated according to the frequency of "weekly cleaning+monthly deep maintenance", with the core of "using the right water and cleaning frequently":

1. Humidification water tank (core water supply source)

  • Daily: Change distilled water or deionized water once a week (avoid using tap water to prevent scale from blocking the humidification nozzle), keep the water level between the "MIN-MAX" scale, and replenish it in a timely manner when it is insufficient.

  • Deep cleaning: Empty the water tank every month, soak it in 5% citric acid solution for 30 minutes, use a soft bristled brush to clean the tank wall (with a focus on cleaning the outlet filter), rinse it with clean water 3 times, air dry and refill with water.

2. Drainage water tank (storing defrosting water/wastewater)

  • Check the water level daily and immediately drain it when it reaches 2/3 of the height to avoid overflowing and corroding the bottom of the equipment.

  • Wipe the inner wall of the water tank with warm water and neutral detergent every month to remove residual dirt and odors, and air dry before returning to its original position.

3、 Equipment basic maintenance: fault prevention

In addition to the core components, the maintenance of the fuselage, pipelines, and control systems is equally critical, and should be carried out according to the "daily inspection+regular calibration":

(1) Air filter screenDisassemble and clean once every 1-2 months, blow off the dust with compressed air (or rinse with clean water to dry), replace the damaged filter screen in a timely manner to prevent dust from blocking the air duct and causing uneven temperature changes.

(2) Sealing strip: Wipe the door sealing strip with a dry cloth every week to remove oil and dust; Apply a thin layer of silicone grease (special food grade) on the surface of the sealing strip every 3 months to prevent aging and cracking, and ensure sealing performance.

(3) Sensors and pipelines: Please have a professional calibrate the temperature sensor every 6 months(Ensure an error of ≤± 0.5 ℃); Check the interface of the cold and hot pipelines. If there is any leakage of condensed water, tighten the sealing components in a timely manner.

(4) Body and heat dissipation: Clean the surface dust of the device daily, wipe it with a damp cloth (do not rinse directly with water); Reserve a heat dissipation space of ≥ 30cm around the device to avoid covering the top or side heat dissipation ports and prevent the controller from overheating.

4、 Core principles of maintenance

1. Power off before operation to avoid short circuits caused by live cleaning; 2. All cleaning tools should be made of soft materials (plastic, cotton cloth) to prevent scratching by hard objects; 3. Establish a maintenance ledger to record defrosting, cleaning, and calibration times for easy traceability; 4. If any abnormalities are found (such as frequent freezing after defrosting or water tank leakage), immediately stop the machine and contact after-sales service. Do not disassemble the machine without authorization.