Phenomena, causes, and solutions for pressure source faults in pressure calibrators
Date: 2025-05-08Read: 13
Pressure calibratorIt is a precision equipment used for calibrating instruments such as pressure transmitters, pressure switches, and pressure gauges. Common faults involve modules such as pressure sources, sensors, electrical signals, and mechanical components. The following are typical fault phenomena, cause analysis, and handling methods:
Fault phenomenon Possible reasons treatment method
Unable to build pressure Air pump/hand pump damaged Replace the pump body or sealing ring
Pipeline leakage (loose joints, aging sealing rings) Use soapy water to detect leaks, tighten joints or replace sealing rings
Valve not closed (e.g. pressure relief valve stuck) Clean or replace valves
Unstable pressure/large fluctuations Gas source contains impurities (oil, water) Install a filter and use a dry and clean air source
Pressure controller PID parameter imbalance Re tune the PID or contact the manufacturer for calibration
Slow boost speed Insufficient pump power Check the power supply voltage and replace the high-power pump
Pressurization of large volume containers (such as gas storage tanks) Segmented pressurization or selection of high-pressure gas source