(1) Measurement range: pH: 0-14.00pH;
Temperature: 0-60 ℃.
(2) Basic error: pH: ± 0.05pH;
Temperature: ± 0.5 ℃.
(3) Automatic temperature compensation range: 0-60 ℃.
(4) Temperature compensation error: ± 0.03pH ± 1 word.
(5) Stability: ± 0.02pH/24h.
(6) Input impedance: ≥ 1 × 1012 Ω.
(7) Input current: ≤ 1 × 10-12A.
(8) Water sample pressure: ≥ 0.6MPa.
(9) Current isolation output: 0-10mA (load resistance<1.5K Ω); 4-20mA (load resistance<750 Ω).
(10) Data storage capacity: 55 hours (1 hour/10 seconds), 14 days (1 hour/1 minute),
68 days (1 hour/5 minutes), 412 days (1 hour/30 minutes).
(11) Data retention time for continuous power failure: 6 years.
(12) Network output: Balanced differential transmission. Distance of 1.2Km; with the addition of a relay, the distance is 10Km long.
(13) Alarm relay normally open contacts: 3A 240VAC, 6A 28VDC or 120VAC.
(14) Power supply: 85-265VAC 50 ± 1Hz.
(15) Power consumption: 10W.
(16) Dimensions of electronic unit: 135 (length) x 135 (width) x 235 (depth) mm;
Installation hole size: 138 × 138mm, with a frame of 144 × 144mm.
(17) Electronic unit mass: 2.5Kg.
(18) Flow type inlet and outlet pipe connection: hard or soft pipe, with an outer diameter of Ø 10 or Ø 6, Ø 8, Ø 12, and Ø 14.
(19) Normal working conditions:
a) Environmental temperature: 0-60 ℃;
b) Relative humidity: ≤ 90%;
c) No vibration, no corrosive gases, no direct sunlight. No magnetic or electric field interference.