Measurement conditions for aerosol capture capacity of atomizing core smoking machine
Date: 2022-03-20Read: 8

Everyone is familiar with the atomizer core, and I don't know when it became popular. Anyway, everyone knows that the atomizer core is an electronic product that imitates cigarettes. Through atomization and other means, nicotine and other substances are turned into steam for users to smoke. Since its inception, it has been labeled as "long-lasting" by businesses and is sought after by many smokers. Below, the editor will introduce to you about the atomizer coreThe main characteristics of smoking machines and the measurement conditions for aerosol capture are welcome to read!

O1CN01Vj6Zv42EUB4x0pi3K_!!2056018747-0-cib.jpg


Atomization corerange hoodThe main characteristics are:

The outer box is made of steel with baking paint, and the inner box is made of stainless steel. It has anti explosion treatment holes and shock-absorbing and pressure relief materials installed in all directions that may be directly impacted. The window glass is tempered with explosion-proof treatment for safe use. The test uses a high-speed booster pump, a proportional valve to accurately control the air supply volume, and real-time feedback data from sensors. Adopting PLC control and touch screen human-machine interface control, the testing time, frequency, intake pressure, etc. can be set, and the operation is convenient and simple.

Atomization coreMeasurement conditions for aerosol capture of smoking machines:

Atomization coreSmoking machine conditions (suction time: 3 ± 0.1s, suction frequency: 30 ± 0.5s, suction capacity: 55 ± 0.3mL)

Test atomization coreThe number of samples is 3 (prepare at least 3 times the number of samples to be aspirated)

If the number of openings is greater than 100, capture 20 openings on the same filter and suction them 5 times in sequence. Take the 1st, 3rd, and 5th filters for analysis;

If the number of openings is greater than 60, capture 20 openings on the same filter and suction them three times in sequence to analyze each filter;

If the number of suction ports is less than 60, collect 10 suction ports on the same filter and suction them 3 times in sequence, and analyze each filter separately;

Thank you for reading. The above is an introduction to the main characteristics of the atomizer core smoking machine and the measurement conditions for aerosol capture. Some of the content is from the internet and is for reference only. If you would like to learn more aboutAtomization coreFor detailed information about smoking machines, please feel free to call us for further inquiries. We are dedicated to serving you.