Split type electromagnetic flowmeterInstallation requirements for straight pipe sections:
Sensors have certain requirements for the upstream and downstream straight pipe sections of the installation point, otherwise it will affect the measurement accuracy.
1. If there is a tapered pipe upstream of the sensor installation point, there should be an equal diameter straight pipe section of not less than 15D upstream of the sensor and an equal diameter straight pipe section of not less than 5D downstream.
2. If there is a gradually expanding pipe upstream of the sensor installation point, there should be an equal diameter straight pipe section of not less than 18D upstream of the sensor, and an equal diameter straight pipe section of not less than 5D downstream.
3. If there is a 90 ° elbow or lower joint upstream of the sensor installation point, there should be an equal diameter straight pipe section of not less than 20D upstream of the sensor, and an equal diameter straight pipe section of not less than 5D downstream.
4. If there are two 90 ° elbows on the same plane upstream of the sensor installation point, there should be an equal diameter straight pipe section of not less than 25D upstream of the sensor and an equal diameter straight pipe section of not less than 5D downstream.
5. If there are two 90 ° elbows on different planes upstream of the sensor installation point, there should be an equal diameter straight pipe section of not less than 40D upstream of the sensor and an equal diameter straight pipe section of not less than 5D downstream.
6. The flow regulating valve or pressure regulating valve should be installed as much as possible outside 5D downstream of the sensor. If it must be installed upstream of the sensor, there should be an equal diameter straight pipe section of not less than 50D upstream and an equal diameter straight pipe section of not less than 5D downstream.
Special attention
1. If a valve is installed near the upstream of the sensor installation point, constantly opening and closing the valve will have a great impact on the service life of the sensor, and it is very easy to cause * damage to the sensor.
2. Sensors should be avoided as much as possible from being installed on very long overhead pipelines. Over time, the sagging of the sensor can easily cause sealing leakage between the sensor and the flange. If installation is necessary, pipeline fastening devices must be installed at 2D positions upstream and downstream of the sensor.