Dear customer, in order to ensure the operational effectiveness of the testing equipment and extend its service life, it is recommended that the daily use of Shanghai Baotian Testing Equipment should meet the following requirements:
1experimentequipmentEnvironmental requirements for use
- Environmental temperature requirements: 5-25 ℃ to ensure good performance, 25-35 ℃ can operate normally (for air-cooled test chambers); Water cooled equipment should try to control the room temperature between 5-25 ℃, and the larger the equipment volume, the lower the temperature should be as much as possible.
- Humidity requirement ≤ 85% RH
- Air pressure requirement: 86KPa~106KPa
- Electromagnetic environment: There is no strong electromagnetic radiation source nearby
- Site drainage: There is a drainage floor drain within 2 meters next to the test chamber
IIexperimentequipmentSafe storage conditions
- When the test chamber is not in use, the storage environment of the test chamber should always be kept above 0 ℃;
- If the storage environment is below 0 ℃, for water-cooled test chambers and humid heat chambers, all remaining water in the equipment should be drained in a timely manner to avoid freezing and damaging the pipelines.
IIIexperimentEquipment usage precautions
- To ensure personal safety, the metal casing of this equipment should be reliably grounded;
- The power supply of the equipment should meet the requirements of the nameplate to ensure that the blowing direction of the condensing fan of the refrigeration unit is forward air and then out air, that is, the outlet air is blown towards the compressor;
- The parameters of the programmable temperature and humidity controller should not be changed arbitrarily except for normal settings, in order to prevent system control confusion and malfunction;
- To prevent system misoperation and reduce control accuracy, try to avoid opening doors during the experiment;
- The lighting inside the box is only for observation purposes and should not be turned on for a long time to avoid affecting control accuracy;
- During equipment operation, duty personnel must be arranged to conduct routine observations of the equipment's operating status.
- Referring to the maintenance requirements for electrical products, professional personnel conduct regular inspections and maintenance of all components of the equipment. The maintenance cycle can be selected from three months, six months, or one year based on the user's own situation to ensure the reliable and safe operation of the equipment components.