The continuous measurement method for DO is the diaphragm electrode method. The diaphragm is composed of polytetrafluoroethylene fibers, polyethylene, etc., with platinum as the positive electrode and aluminum and lead as the negative electrodes. The two metal electrodes are immersed in an electrolyte solution, and the electrodes and electrolyte solution are housed in a small chamber of an aerobic semi permeable membrane. Connect the current measurement circuit, and molecular oxygen in the water diffuses through the membrane to the electrode surface, causing an electrode reaction. When the electrode parameters are constant, the stable diffusion current is proportional to the oxygen concentration in the water sample at a certain temperature.
Beijing Times New Wei Portable Micro Dissolved Oxygen Analyzer
The TP350 portable trace dissolved oxygen analyzer is a highly functional and easy-to-use high-precision microgram level dissolved oxygen analyzer. Chinese display and menu guidance make operation extremely convenient, with features such as multiple functions, high measurement accuracy, and easy portability. Can be used for measuring trace dissolved oxygen in solutions from various industries.
Beijing Times New Wei Portable Micro Dissolved Oxygen Analyzer
lHigh precision measuring electrode, imported membrane head with short response time, stable measurement, and high temperature resistance.
lAdopting advanced surface mount technology and integrated design, the high integration circuit design is stable and durable.
lAdvanced microcontroller technology, high performance, and low power consumption.
l24 bit A/D signal acquisition, high measurement accuracy, accuracy and reliability.
lChinese menu style operation, easy to understand, quick and convenient to operate.
lThe data loop storage function automatically clears overflow data, with simple operation and convenient query. It can store data for more than 10 years after power failure.
lLong continuous use time, up to 8 hours of measurement after fully charged.
lCompact and beautiful in appearance, easy to carry.
Protection level: IP65
Temperature sensor: Pt1000
Power supply: AC (85-265) V
Frequency (45-65) Hz
Power: ≤ 1W
Dimensions: 285mm × 60mm × 255mm
Weight: 2.28kg
Display: 128 × 64 LCD, Chinese display
Measurement range: (0.0-100) μ g/L, (100-1000)
μg/L、(1.00~20.00) mg/L(自
Dynamic switching)
Resolution: 0.1 μ g/L (0.0-100) μ g/L
1.0μg/L (100 ~1000)μg/L
0.01mg/L (1.00~20.00)mg/L
Repeatability: ≤ 1%
Accuracy: ± 1.0% FS
Response time: T90<5 minutes (25 ℃)
Temperature compensation range: (0-60.0) ℃
Water sample temperature: (5-45) ℃
Environmental temperature: (5-45) ℃
Environmental humidity: ≤ 90% RH (no condensation)
lAccessories Guide
lDissolved oxygen membrane head
lElectrode filling solution
lDissolved oxygen electrode
Precautions
1. The entire measurement system has a long response time, membrane rupture, and no oxygen medium
If the current in the medium increases, it is necessary to replace the membrane or electrolyte. each
Re polarization and calibration are required after membrane or electrolyte replacement.
2. Place the electrode into a flow cup, which can be obtained for on-site testing at any time
Quantity. The circulation cup allows water to flow from the bottom and water to flow from the top.
When measuring low oxygen, it is recommended to use a hard tube to connect the electrode flow cell,
To prevent oxygen from seeping into the waterway through the hose. Inlet water sample
Keep the flow rate as constant as possible.