The CM-230D conductivity meter is an upgraded product of our company's CM-230 conductivity monitoring instrument. On the basis of maintaining the original product performance, it uses a microcontroller digital circuit for measurement and temperature compensation. This product is widely used for online conductivity measurement and monitoring of water treatment equipment such as seawater desalination, and can meet the electrostatic interference and electrode chuck connection requirements of stainless steel equipment. It has the characteristics of more convenient use, easier operation, and more complete performance.
● Equipped with automatic range switching;
● Equipped with automatic zero adjustment;
● It has the ability to switch display temperature values;
● Equipped with electrode constant panel input;
● Equipped with conductivity upper limit alarm setting and control contact output (optional upper and lower limit alarm output);
● Equipped with alarm control delay setting;
Measurement range:0.00~199.9mS/cm; Among them, 0.00~19.99mS/cm and 0.00~199.9mS/cm have automatic range switching.
Temperature compensation:Equipped with automatic temperature compensation, with a compensation range of 0-100 ℃, a compensation benchmark of 25 ℃, and a compensation coefficient of 2.0%/℃; Temperature compensation component NTC10K Ω.
Display mode:31/2Direct reading display of conductivity value on 0.56 "LCD (liquid crystal with backlight);
Instrument level:Level 2.0;
Electrode configuration:K=10.00 or K=1.00 conductive electrode, G1/2 or G3/4 pipe type threaded connection or 50.5 chuck connection, wire length of 5m;
| Electrode configuration | measurement range | connection method |
| K=1.00 | 0~19.99mS/cm | 4 'thread or sanitary grade chuck |
| K=10.0 | 0〜199.9mS/cm | 6 'thread or sanitary grade chuck |
Alarm control and signal output:This instrument has a conductivity upper limit alarm setting and alarm contact (switch quantity) output, with a contact capacity of 3A/220VAC;
Current signal output:Equipped with isolated 4-20mA current signal output, the 20mA operating value can be set arbitrarily, and the output load is 750 Ω; (Active)
Power supply:AC220V ± 10%, 50Hz; (or DC24V);
Instrument size:
Frame size: 96 × 48 (mm)
Hole size: 91 × 45 (mm)
Instrument depth: 100 (mm)
Product standards:Q/TDCB04;