CS1005 High Current Electrochemical Workstation

NegotiableUpdate on 08/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

CS1005 High Current Electrochemical Workstation Compared with CS series single channel, CS1005 High Current Electrochemical Workstation has increased current output and measurement range, and the maximum output current can reach 5A

Product Details

Hardware parameter indicators

Constant potential control range ±10V
Constant current control range ±5.0A
Precision of potential control 0.1% x full-scale reading ± 1mV
Current control accuracy 0.1% x full-scale reading
Potential sensitivity 10μV(>100Hz), 3μV(<10Hz)
current sensitivity <100pA
Potential rise time <1μS(<10mA),<10μS(<2A),<20μS(>2A)
current range 20A~2μA, 8 levels in total
Reference electrode input impedance 1012Ω||20pF
maximum output current 5.0A
Slot pressure output ±12.5V
Current scanning increment 1mA @1A/mS
CV and LSV scanning speed 0.001mV~10000V/s
Potential scan potential increment 0.076mV @1V/mS
CA and CC pulse widths 0.0001~65000s
DPV and NPV pulse widths 0.0001~1000s
SWV frequency 0.001~100KHz
Minimum potential increment of CV 0.075mV
AD data collection 16bit@1MHz , 20bit @1kHz
Current and potential range Automatic setting
DA resolution 16bit
Establishment time 1μS
Low Pass Filter 8-segment programmable
Communication interface USB2.0
Instrument weight 6.5Kg
Dimensions (cm) 36.5(W)X30.5 (D)X16 (H)

Electrochemical impedance measurement indicators

signal generator
frequency response 10μHz~115KHz
Amplitude of communication signal 1mV~2500mV
frequency accuracy 0.005%
DC bias voltage -10V~+10V
DDS output impedance 50Ω
wave form Sine wave, triangular wave, square wave
Sine wave distortion rate <1%
Scanning method Logarithmic/linear, increasing/decreasing
Signal Analyzer
Maximum integration time 106One cycle or 105S
Measurement time delay 0~105second
Minimum integration time 10mS or the maximum duration of one cycle
DC bias compensation
Potential compensation range -10V~+10V
Current compensation range -1A~+1A
Bandwidth adjustment Automatic or manual setting, with a total of 8 adjustable levels