1Usage Overview:
Drug shaker (penicillin shaker) is an indispensable instrument for clinical analysis of various blood coagulation tests, passive blood cell agglutination tests, complement binding tests, latex agglutination tests, etc. This instrument can be used by health departments and biochemical laboratories of scientific research institutions for the analysis of trace samples, biological bacteria, virus immunity, clinical testing, and routine laboratory tests.
2、 Technical performance (including data and specifications):
1. Oscillation frequency range: minimum25-35(HZ)Maximum45-60(HZ)
2Vibration acceleration:XAxis (left-right direction)0.5-1.3g
YAxis (front to back direction)0.5-2g
ZAxis (vertical direction)0.25-0.5g
3. Power supply voltage: AC220V±10V,50±2HZ
4Power consumption: <10VA
5Volume:210x165x190(mm)
3、 Usage:
This instrument is fixed with an eccentric heavy hammer at the shaft end of the micro motor. When the motor rotates, it generates an eccentric torque, which reacts to the hoop shaped rubber suspension, causing it to vibrate and form a displacement of up and down, left and right, and front and back movements, forming a periodic motion trajectory of a closed spiral line. This causes the material points reflected in the micro holes of the concave glass plate of the tray to continuously move in the direction of the vibration trajectory, achieving uniform mixing. The DC motor of this instrument adopts silicon speed regulation, with a simple and reliable circuit. This instrument is easy to operate, highly sensitive, requires less medication and blood, and has a fast detection speed.
4、 Attention:
1. The oscillation table is supported by steel wires and should not be heavily pressed.
2. Before use, please remove the fixed components inserted at the four corners.
3. For safety reasons, the power supply must be grounded.
4. To prevent movement, please fill the double-sided tape between the four feet and the desktop