host |
|
|
operation temperature |
-5°- 50°C |
|
Storage temperature |
-20-80 ℃ (pH, pH/ORP sensor storage temperature is 0-60 ℃) |
|
Maximum bearing depth |
225 meters |
|
communication |
Computer interface: Bluetooth wireless communication RS-485、USB; |
|
data storage |
225000 sets of data |
|
sampling frequency |
1HZ |
|
Software operating system |
Supports all Windows software and Android system |
|
Archer on-site display recorder |
|
|
operating system |
Windows mobile 6.5.3 |
|
size |
Width 12cm, height 24cm |
|
protection grade |
IP68 |
|
operation temperature |
-10°- 50°C |
|
Storage temperature |
-20 - 80℃ |
|
data storage |
512M, Scalable up to 8GB |
Technical parameters of sensors
|
parameter |
measuring principle |
range |
accuracy |
resolution |
|
temperature |
Thermistor method |
-5°- 50°C |
±0.10°C |
0.01°C |
|
dissolved oxygen |
fluorescence |
0 -50mg/L |
0- 20mg/L ± 0.1 mg/L |
0.01 mg/L |
|
20-30mg/L ±0.15 mg/L |
0.01 mg/L |
|||
|
30-50mg/L 5% of reading |
0.1 mg/L |
|||
|
0-500% air saturation |
± 1% of the reading |
0.1% air saturation |
||
|
conductivity |
Graphite electrode method |
0 to 275 mS/cm |
When 0 to 10 mS/cm |
0.0001 - 0.01 mS/cm |
|
When 10-100 mS/cm |
||||
|
At 100-275 mS/cm |
||||
|
conductivity |
Graphite electrode method |
0-275 mS/cm |
± 0.5% of reading or |
0.0001-0.01 mS/cm |
|
salinity |
Calculated by conductivity and temperature |
0—70 pss |
0.2 pss |
0.01pss |
|
Total Dissolved Solids (TDS) |
Calculated by conductivity and temperature |
0—65 g/L |
± 5% of the reading |
0.1 |
|
pH |
Glass composite electrode method |
0 - 14units |
± 0.1, within the calibration temperature range of ± 10 ℃ |
0.01 units |
|
ORP |
Platinum electrode method |
-999-999mV |
±20mV |
1mV |
|
turbidity |
90 ° scattering method |
0-5000FNU |
0-400 FNU |
0-999 FNU |
|
400-5000FNU |
1000-5000FNU |
|||
|
depth |
Pressure sensor |
0-10m |
0.003m |
0.001m |
|
0 - 25m |
0.05m |
0.01m |
||
|
0-200m |
0.4m |
0.01m |
||
|
chlorophyllA |
fluorescence |
0 -500μg/l |
± 3% of full scale |
0.01μg/l |
|
blue green alga |
fluorescence |
0-40000 μ g/l or |
± 3% of full scale |
0.01μg/l |
|
blue green alga |
fluorescence |
0-750μg/l |
± 3% of full scale |
0.01μg/l |
|
Ammonia nitrogen |
Ion selective electrode method |
0 - 100mg/L-N |
± 5% or ± 2 mg/L |
0. 1mg/L-N |
|
nitrate |
Ion selective electrode method |
0 - 100mg/L-N |
± 5% or ± 2 mg/L |
0. 1mg/L |
|
chloride |
Ion selective electrode method |
0 - 18000mg/L |
± 5% or ± 2 mg/L |
0. 1mg/L |
|
sodium ion |
Ion selective electrode method |
0 - 20,000 mg/l |
± 5% or ± 2 mg/L |
0. 1mg/L |
|
calcium ion |
Ion selective electrode method |
0 - 40,000 mg/l |
± 5% or ± 2 mg/L |
0. 1mg/L |
|
bromide |
Ion selective electrode method |
0 - 80,000 mg/l |
± 5% or ± 2 mg/L |
0. 1mg/L |
|
TDG |
Pressure method |
400-1400mmHg |
±1% |
0.1mmHg |
|
rhodamine |
fluorescence |
0-1000ppb or |
± 3% of full scale |
0.01ppb |
|
fluorescein |
fluorescence |
0-1000ppb or |
± 3% of full scale |
0.01ppb |
|
CDOM/FDOM |
fluorescence |
0-1250 ppb (QS) or |
± 3% of full scale |
0.4 ppb |
|
crude |
fluorescence |
0-1,500 ppb |
± 3% of full scale |
0.2 ppb |
|
Refined oil |
fluorescence |
0-10,000 ppb |
± 3% of full scale |
2 ppb |
|
Environmental light PAR |
fluorescence |
0 - 10,000μmol/sm2 |
+/-5%, or |
1μmol/sm2 |