Overview

Product Description: According to the Particle Measurement Program (PMP), calculating the number of particles emitted by an engine is valuable in engine emission research and is a key component of engine type certification

Product Details

发动机排气凝聚核粒子计数器 3790

Product Description: How to Translate

According to the Particle Measurement Program (PMP), calculating the number of particles emitted by an engine is valuable in engine emission research and is a key component of engine type certification. The TSI Engine Exhaust Condensation Particle Counter (EECPC) model 3790A is designed specifically for these needs and meets the requirements of UN ECE R83. It is a valuable tool for any customer who needs to measure particles in engine exhaust.

Details

The engine exhaust condensed core particle counter 3790A (EECPC) meets all the requirements of PMP for Euro 5 and Euro 6 regulations. This instrument is built on the proven TSI second-generation alcohol based condensed core particle counter technology, providing users with robust, reliable, and highly reproducible performance indicators. The engine exhaust condensation nucleus particle counter 3790A adopts new improvements, such as anti overflow design, condensate discharge system, removable saturator, USB communication interface, etc. 3790A has a detection limit lower limit with a particle size greater than 20nm, thus meeting the PMP regulation's requirement to eliminate volatile and semi volatile nanoparticle components formed by the nucleation of gaseous compounds.

TSIThe company's 379020A rotary disc diluter and 379030 hot air conditioner can be connected to EECPC to eliminate nanoparticles formed by the condensation of volatile components in engine exhaust emissions, as proposed in the GRPE/PMP regulation draft.

Product application:

ŸEmission testing of gasoline and diesel engines

Product features and advantages:

ŸMeet the lower particle size detection limit of PMP and provide counting efficiencies of D10, D25, D50, and D90

ŸExcellent linear response with a linear correlation coefficient greater than 0.95 for particle concentrations ranging from 1 to 10000 particles per cubic centimeter

ŸCompared to traceable standards, counting accuracy can vary within ± 10%

ŸSingle particle counting under full flow conditions

ŸCalibration fully complies with PMP specification requirements

ŸReaching 0.1 particles per cubic centimeter, readability

ŸInternal pulse amplitude display shows measurement quality

Ÿ10 HzModal analysis of data rate