High temperature aging testing machine, also known as high and low temperature testing chamber, high and low temperature aging testing machine, high and low temperature testing chamber, etc., is suitable for reliability testing of industrial electronic products at high and low temperatures. Suitable for: electrical and electronic, aerospace, optoelectronics, hardware, plastics, scientific research, and more. Satisfy with national and military standards.
Create Chinese quality instruments with the spirit of craftsmanship




☆The box is processed and formed by CNC machine tools, with a beautiful and elegant appearance, exquisite craftsmanship, and simple operation;
☆Large observation window and adjustable sample storage rack;
☆Test holes can be opened on both sides of the box for external testing of power and signal lines to prevent mutual interference;
☆Manual water replenishment for humidity testing/Automatic water replenishment can be configured and automatically recycled for use;
☆Complete communication protocol, supportingUSB、RS-232、R-485、RJ-45Convenient communication methods for remote monitoring and data collection;
☆【Refrigeration unit based on French TaikangTecumsehA low-noise air-cooled unit is composed of a fully enclosed compressor, which uses environmentally friendly refrigerants, three-dimensional pipeline vibration reduction design, nitrogen protection without oxidation welding, regulation of refrigeration and dehumidification capacity, gas-liquid bypass regulation, evaporation temperature regulation and other comprehensive technologies to ensure efficient and reliable operation of the refrigeration unit.

High and low temperature aging testing machine

Back view of high and low temperature aging testing machine

Left side diagram of high and low temperature aging testing machine

Right side diagram of high and low temperature aging testing machine
③ Open the equipment distribution panel to check for loose components and circuit wiring.
④ If the use is stopped for a long time, the high and low temperature test chamber should be powered on and lit regularly for 1 hour.
If you encounter any problems during use, you can contact our customer service in a timely manner and provide free technical support. In addition, we also offer repair services.
| PRODUCT MODEL | LS-TH/LS-GDH | ||||||||||||||
| Inner box capacity | L | 30 | 80 | 100 | 150 | 225 | 408 | 800 | 1000 | ||||||
| Inner box size (mm) | wide | 300 | 400 | 500 | 500 | 600 | 750 | 800 | 1000 | ||||||
| tall | 400 | 500 | 500 | 600 | 750 | 850 | 1000 | 1000 | |||||||
| deep | 250 | 400 | 400 | 500 | 500 | 650 | 1000 | 1000 | |||||||
| Outer box size (mm) | wide | 460 | 650 | 750 | 750 | 850 | 980 | 1250 | 1250 | ||||||
| tall | 1000 | 1540 | 1540 | 1640 | 1810 | 1890 | 2050 | 2050 | |||||||
| deep | 900 | 1200 | 1350 | 1450 | 1450 | 1580 | 1750 | 1950 | |||||||
| Product weight | kg | 128 | 190 | 210 | 240 | 280 | 320 | 480 | 550 | ||||||
| Low temperature range | R series: 0 ℃ H series: -20 ℃ Z series: -40 ℃ P series: -60 ℃ S series: -70 ℃ | ||||||||||||||
| High temperature range | +150℃ | ||||||||||||||
| Humidity range | 20%~98%RH | ||||||||||||||
| temperature fluctuation | ±0.5℃ | ||||||||||||||
| Temperature deviation | ±2.0℃ | ||||||||||||||
| Humidity deviation | Above 75% RH ≤+2/-3% R.H, below 75% RH ≤ ± 5.0% RH | ||||||||||||||
| Temperature and humidity range chart |
|
||||||||||||||
| Heating rate (min) | About 3.5 ℃/min Nonlinear no-load room temperature RT~+150 ℃35 min | ||||||||||||||
| Cooling rate (min) | About 1.2 ℃/min Nonlinear no-load room temperature RT~-40 ℃60 min | ||||||||||||||
| Inner box material | SUS304 stainless steel | ||||||||||||||
| External material | Cold rolled steel plate electrostatic spraying surface treatment or stainless steel | ||||||||||||||
| Insulation material and thickness | (Thickness 100mm) high-density polyurethane foam | ||||||||||||||
| Cooling method | Air cooled (optional water-cooled). | ||||||||||||||
| refrigeration system | Original imported French Taikang fully enclosed compressor or German Bitzer semi enclosed compressor. | ||||||||||||||
| Protector | Non fuse switch, compressor overpressure, overheating, overcurrent protection, over temperature protection, fan overload protection, fuse, dry burn protector, water shortage protection | ||||||||||||||
| controller | 7-inch touch screen controller, Chinese menu, constant value/program mode operation | ||||||||||||||
| Standard accessories | 1 insulated glass window, 1 lead hole (Φ 50mm hole), 2 sets of sample racks, 1 set of lighting fixtures, and 1 pack of dry and wet bulb gauze | ||||||||||||||
| Power | KW | 3.8 | 4.0 | 4.5 | 5.0 | 6.8 | 7.5 | 10.5 | 12.5 | ||||||
| Environmental temperature for use | -5℃~+35℃ | ||||||||||||||
| Power Supply | AC220V±10% 50/60Hz AC380V±10% 50/60Hz | ||||||||||||||
| Purchase instructions | LS stands for Laisi, TH - Damp Heat Chamber, GDH - High and Low Temperature Chamber. All box sizes inside the test chamber can be customized according to customer needs. | ||||||||||||||
Knowledge of maintenance and upkeep for high and low temperature experimental chambers
① Check whether the equipment placement environment meets the environmental temperature and ventilation requirements of -5 ℃~35 ℃.
② Check if there is dust accumulation on the condenser. Use a vacuum cleaner or high-pressure air to clean the dust on the condenser.
③ Open the equipment distribution panel to check for loose components and circuit wiring.
④ If the use is stopped for a long time, the high and low temperature test chamber should be powered on and lit regularly for 1 hour.
If you encounter any problems during use, you can contact our customer service in a timely manner and provide free technical support. In addition, we also offer repair services.