The SS-1000R imaging glare brightness meter is produced by Hangzhou Dual Color Intelligent Detection Instrument Co., Ltd. It uses a 16 million pixel 4/3 inch COMS, pixel size 3.45umx3.45um, matched with a human visual function V (λ) color filter, and has a measurement accuracy of level one. It can measure multiple parameters such as Unified Glare Index (UGR), meet multiple international and domestic standards, and is widely used in various fields such as light sources and lighting. It comes with a two-year warranty and lifetime maintenance.
Product Details

Imaging glare brightness metertechnical indicators

SS-1000RImaging glare brightness meterThe measurement system adopts a high-precision imaging brightness meter,Unified glare index has been achieved through precision electronic control pan tilt and dedicated measurement and control softwareUGROutdoor sports stadium glare valueGRIncremental threshold for road lightingT1Window glare valueDGIHigh precision measurement and analysis,satisfyCIE 115TheGB/T 5700-2023TheGB/T 50034-2024TheJGJ/T163-2008TheJGJ/T153-2007TheCJJ45-2015Waiting for standard requirements

Performance parameters:pixel size3.45umx3.45um CMOSsize4/3Inch image sensor and high-precision matchingV(λ)filter;Effective pixels1600Ten thousand, resolution@frame rate4112x4112maximum@23.5fpsADwidth12bit Point matrix,Can capture the entire scene:Simultaneously measure the brightness of each pixel point.

Pixel Capture:Automatic, point, multi-point, multi-point connection, circle, square, polygon, ellipse, curve, etc

Test parameters: Unified glare valueURGGlare index(TI)The glare value of sports venues(GR)The discomfort and glare of the windowDGIBrightness, brightness uniformity, maximum value, minimum value, average value, brightness distribution curve, brightness pseudo color

Key Features:

1High precision measurement: strict matchingHuman visual functionV(λ)The color filterCan accurately obtain real and accurate brightness data.

2Rich software features: It has functions such as single point, multi-point, and area brightness and chromaticity uniformity distribution pseudo color maps, and each parameter can be set with upper and lower limits for judgment.

3Exposure area: You can set the exposure area to move any pixel size

4Support secondary calibration function

5Rich interfaces: equipped with input and outputIOInterface, compatible with customersPLCWait for automated equipment to trigger testing and output qualified or unqualified results(PASS/NG)Instructions can also be provided for third-party developmentSDKSoftware package.

Main applications:Main applications: testing of brightness in various fields such as point light sources, surface light sources, fire emergency lights, navigation lights, urban lighting, tunnel lights, automotive ambient lights, road brightness, landscape lighting, dashboard brightness, display screen brightness, light sources and light-emitting devices, film and television, traffic signal signs, architecture, and atmospheric lightingLEDTesting of finished lighting fixturesLEDUniformity testing of lighting fixtures, etc.

CIE 117-1995Indoor lighting discomfort and glareCIE 112-1994Glare evaluation system for outdoor sports venues and area lightingCIE 140-2000Road lighting calculation

GB50034-2013Architectural Lighting Design StandardsCJ145-2015Design Standards for Urban Road LightingGB 7793-2010Lighting and Hygiene Standards for Primary and Secondary School ClassroomsJG1153-2016Sports venue lighting design and testing standards, other glare related standards

GB4599-2007TheQC/T413Test standards for automotive ambient lightingFire emergency light New National StandardGB17945-2010for automotive use LEDheadlightGB 25991-2010

threePerformance indicators

parameter name

parameter indicators

optics

4/3'

Sensor Model

COMS

camera type

Ycamera

pixel size

3.45umx3.45um

Effective pixels

1600ten thousand

exposure method

Frame exposure

resolution@frame rate

4112x4112maximum@23.5fps

signal-to-noise ratio

40dB

dynamic range

73.6dB

sensitivity

1 146mV

ADwidth

12bit

Output pixel width

12bit

Collection Mode

continuous/soft trigger/Hardware trigger

Maximum gain(multiple)

250

Exposure area

Can be set to move any pixel size

Exposure time range(ms)

0.0141~29546.9

frame buffer

128MB

lens mount

OptionalCSC

data interface

USB3.0

power supply

5V

power

<2W

Overall dimensions

155*131*126mm

weight

500g

measurement range

0.01-10ten thousandcd/m²

measurement accuracy

± 4% reading ± 1 reading) under light source A

Measurement repeatability

±0.1%

focal length

12mm

sensor type

Human visual functionV(λ)Color filter matching