E-mail
2197115204@qq.com
-
Phone
13761613451
-
Address
No. 41, Lane 2688, Yindu Road, Minhang District, Shanghai
Shanghai Heyi Instrument Co., Ltd
2197115204@qq.com
13761613451
No. 41, Lane 2688, Yindu Road, Minhang District, Shanghai
iodine-131Automatic nuclide packaging instrumentIntegrated with pin insertion, automatic packaging, activity measurement, intelligent dispensing, visual intercom, intelligent terminal and other functional modules, it can be embedded and installed in a fume hood. Through packaging management software, it can be remotely controlled and operated.The intelligent terminal provides medication patient calling and dosage information display, and real-time visual voice intercom.
2. configuration
2.1Automatic nuclide packaging instrument1Taiwan;
2.2Pin insertion device1Set;
2.3Dispensing device1Taiwan;
2.4 Packaging management software and computer1Set;
2.5Intelligent terminal device1Taiwan;
2.6 Instruction manual and certificate of conformity1Share;
3.parameter
3.1Packaging method: Installed in a shielded fume hood, suitable for external independent activity meters, can be freely removed for measurement calibration, seamless compatibilityCRC-55tROr domestically produced activity meters;
※3.2Intelligent terminal: Large display screen, providing intelligent patient calling and voice broadcasting, guiding patients to complete the process of self pickup and medication taking, and displaying the current patient's personal information and medication activity information for patient confirmation, avoiding the occurrence of medication errors;
※3.3Medication dispensing: Personal medication is dispensed into a medication cup, which can be remotely adjusted for dispensing position and features splash proof design, oral medication cup sensing, and high-definition camera;
3.4 Activity range of packaged drugs:1mCi-10000mCi.
※3.5 Automatic transfer and dilution of mother liquor, automatic measurement of specific activity and real-time display, and manual addition of diluent;
※3.6Equipped with dual cameras for real-time display of medication location, room and oral cup status, video monitoring of patient medication process, and voice communication with patients.
3.7.Real time online measurement of the activity of the original solution and the activity of the packaged sample drug solution to ensure the accuracy of the packaged activity;
3.8Real time display of remaining liquid activity, facilitating the planning of patient prescription activity;
3.9Activity error of packaged samples: ≤±5%;
3.10Minimum packaging volume: ≤0.2ml;
3.11Packaging time: ≤120 second/People;
3.12 Packaging management software: throughTCP/IPNetwork communication, not limited by distance, can remotely control the automatic dilution, transfer, packaging, and dispensing of radioactive drugs within the local area network;
3.13Working voltage:AC220V 50Hz;