The correct adjustment of Kohler lighting system is one of the main tasks of microscope calibration, which is to adjust the illumination light path system, and the key is to adjust the Kohler lighting system

Product Details

Correct adjustment of Kohler lighting system
One of the main tasks for the correct calibration of a microscope is to adjust the illumination light path system, and the key is to adjust the Kuller illumination system. For every person who uses a microscope, especially those who take microscopic images, they should have a certain understanding and mastery of the principle and adjustment steps of the Kuller illumination system in order to fully utilize the functions of the microscope, and the photos taken can be consistent and complete in effect.
This article is written by a polarizing microscope2px '>Provided throughout the entire area
  

The principle of the Kuller illumination system is simply that the light emitted from any point on the light source emitter can illuminate the field of view of the microscope, and the light emitted from each point on the light source emitter is collected to achieve very sufficient and uniform illumination in the field of view of the microscope.

The purpose of adjusting the Kuller lighting system is to provide uniform and sufficient illumination for the observed field of view, prevent stray light from affecting or interfering with the image system, and prevent the formation of fog on the film during photography. The necessary components of the high adjustment Kuller lighting system include a field of view aperture and a spotlight system that can be adjusted for axis alignment.
select10Objective lens and10X eyepiece;
Place the front lens of the spotlight into the optical path, adjust the aperture stop to a moderate position (not too big or too small), then raise the spotlight to the top position, and adjust the spotlight dial to the bright field of view“J”Location;