LX-D-2 D-type hardness tester for plastic product hardness testing

NegotiableUpdate on 05/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The LX-D-2 plastic Shore hardness tester is a dual pointer D-type Shore hardness tester that can be used for handheld testing or for testing on a separate testing bench of the same model. The LX-D-2 plastic Shore hardness tester is used to determine the hardness values of plastic products, vulcanized rubber, plastic test pieces, and other products. It is designed and produced according to various international and national standards. When the LX-D-2 plastic Shore hardness tester is used to test plastic specimens, it should be tested in accordance with the relevant provisions of GB/T2918-1998 "Standard Environment for Plastic Specimen Condition Adjustment and Testing", and the recommended measurement range is 20-90HD.
Product Details
The LX-D-2 plastic Shore hardness tester consists of a test head, dial, peak pointer return knob, peak pointer (red), locking screw, tracking pointer (black), and other components

Pre test inspection method

1. Before use, check whether the pointer of the LX-D-2 plastic Shore hardness tester points to zero in its free state (if the pointer deviates slightly from zero, you can loosen the upper right corner compression screw, rotate the dial to align with zero, and then tighten the compression screw); If the peak pointer deviates from zero, the peak pointer return knob should be turned first to place the peak pointer close to the right side of the tracking pointer.

2. When the LX-D-2 plastic Shore hardness tester is pressed onto a reference plane to check the zero position, the hard indenter is easily pressed into the reference plane under high load, making it difficult to accurately read. Therefore, a standard measuring block must be placed on the reference plane. When operating, hold the LX-D-2 plastic Shore hardness tester in hand, align the pressure needle with the center hole of Lingkuai, and press down the plastic Shore hardness tester. When the LX-D-2 hardness tester is pressed against the dedicated measuring block and reference surface, the hardness tester should display 10 ± 0.5HD (the thickness of the measuring block is 2.25 ± 0.003mm, which is equal to the depth of the hardness tester when pressed into 10HD). If it cannot be accurately displayed, the hardness tester cannot be used.

Specification parametersSpecs

model LX-D-2
Needle stroke range 0-2.5mm
Dial value 0-100HD
Recommended measurement range 20-90HD
measurement error ±1HD
Pressure at the end of the pressure needle F=445HD(mN);HDFor the hardness reading of Shore D hardness tester
weight 0.16kg
Needle head size SR0.1mm
External dimensions 112×60×25mm