The NK-102E explosion-proof oxygen analyzer complies with the relevant provisions of GB3836.1-2000 and GB3836.4-2000 standards, and has been inspected and qualified by the Explosion proof Electrical Product Quality Supervision and Inspection Center. The explosion-proof mark is Exia Ⅱ CT4
NK-102E Hydrogen Oxygen AnalyzerTechnical Specifications:
1. Measurement range: 0.00-2.00% O2 or 0.00-25.0% O2 (range can be customized).
2. Accuracy: ≤± 1% F.S.
3. Repeatability: ≤± 0.5%.
4. Resolution: 0.01%.
5. Response time: T90 ≤ 10 seconds.
6. Sample gas pressure: 0.05 MPa ≤ inlet pressure ≤ 0.1 MPa. (The outlet must be at atmospheric pressure)
7. Sample gas flow rate: 400 ± 10 mL/min.
8. Output signal: 4-20mA (two-wire system).
9. Working power supply: DC24V.
10. Working environment: Operating temperature: -5 ℃ to+45 ℃.
11. Operating humidity: ≤ 90% RH (no condensation).
12. Dimensions: Transmitter: 140mm (width) x 140mm (height) x 80mm (depth).
13. Explosion proof mark: Exia Ⅱ CT4.
14. Weight: More than 1kg.
15. External safety barrier parameters: Vo ≤ 28VDC, Io ≤ 93mA, Cp ≤ 0.02uF, Lp ≤ 1mH.

NK-102E Hydrogen Oxygen AnalyzerApplication fields:
Online detection and analysis of oxygen content in industrial production processes such as fertilizers, petrochemicals, hydrogen production from electrolyzed water, and biological fermentation for explosion-proof and non explosion proof applications.
Instrument features:
1. Imported electrochemical sensors have the characteristics of long lifespan, high anti-interference ability, high precision, and fast response.
2. On site LCD display, easy to operate for maintenance and calibration, ensuring long-term measurement accuracy.
3. Small size, light weight, easy to install.
4. High reliability intrinsic safety explosion-proof design, IP65 protection level.
5. The primary meter adopts a standard 4-20mA two-wire signal output, with a long transmission distance.