E-mail
whhshjd@163.com
- Phone
-
Address
Wuhan Economic Development Zone Economic Development Zone Wanda Office Building B4-506
Wuhan Haosiheng Electromechanical Co., Ltd
whhshjd@163.com
Wuhan Economic Development Zone Economic Development Zone Wanda Office Building B4-506
DeltaClassicThe unparalleled unique performance of DE-2000
1. Adopting a completely redesigned radiation tube, no high-voltage power lines, no RF noise, and better X-ray shielding.
2. The structure is more precise, shortening the distance between the radiation tube, detector, and the measured sample, and improving the signal by about 40% for certain applications
More than one-third of the machine body adopts an aluminum alloy shell design, and there is a dedicated slot type heat dissipation device on the top of the instrument. The entire system makes heat dissipation very effective, prolongs machine life, and makes X-ray analyzer work more stably, resulting in an extremely low failure rate.
4. The built-in barometer can correct air pressure, and the pressure sensing of the barometer can be adjusted according to different pressures, resulting in more accurate detection values. The instrument is preset at sea level, and customers can test at different altitudes without the need for recalibration.
5. The built-in accelerator can detect motion and vibration, allowing the instrument to stand by when not in use and work when picked up.
6. The intelligent docking station can charge additional batteries and the built-in battery of the instrument simultaneously, and display the charging progress. The docking station can connect to a computer to exchange data, allowing the instrument to be standardized in real time and in standby mode at any time.
7. Even when the instrument is turned on, the battery can be replaced without shutting down. When the battery is unplugged during system operation, there is more than 90 seconds to insert a new battery.
8. The instrument does not require standardization when turned on, and can be tested directly. Standardization is only an option.
9. The instrument casing is made of aluminum alloy and plastic design, which is sturdy and durable. The handle is designed with soft rubber, which has a good hand feel and is more suitable for long-term use.
10. The instrument has good balance and can stand on the workbench during testing without the need for manual support. With a one click button design, it can operate for a long time without fatigue.
11. Industrial grade touchable display screen with integrated sealing design with the host, capable of withstanding harsh outdoor working environments. Sealed instruments can work normally for a long time in rainy and dusty mining environments.
12. The window explosion-proof film design can directly test any shape of object, irregular objects such as shavings and steel wires.
DeltaClassicTechnical performance of DE-2000 alloy analyzer
1. Truly achieve non-destructive, fast, and accurate detection on site, directly displaying the percentage content of elements.
2. Simply place the alloy analyzer in direct contact with the surface of the alloy to be tested, and the alloy grade can be determined on site without waiting or spending time.
The object of the tested sample can be alloy blocks, alloy sheets, alloy wires, alloy slag, or alloy powder.
4. Compensatory testing methods for irregular or small samples can detect very small or few samples, such as filaments with a diameter of 0.04mm that can be immediately identified.
5. The pistol style extendable probe design can detect the interior of pipelines, welds, and other locations.
6. Can detect high temperature materials up to 450 ℃.
7. Alloy grades can be added, edited, and deleted on-site in the library.
8. The world's fastest analysis speed can identify alloy elements in just 2 seconds.
9. The customized Windows CE 6.0 system driver microcomputer display system enables all functions to be completed on-site, while the customized Windows CE only retains the most basic performance related to Windows and Delta systems, making the program more flexible.
10. Without the need for a computer, users can freely specify, view, and zoom in on the spectra of relevant elements on site.
11. Dustproof, anti fog, and waterproof: The all-in-one machine design, with convex and groove structures for soft rubber and plastic components, enables the instrument to have excellent three proof performance, and can withstand harsh working environments such as heavy fog, rain, and dusty work sites.
12. Higher detection accuracy, the average value statistical function of multiple tests can effectively improve the detection accuracy of the instrument.
13. Large icon display, menu driven, and microcomputer WINDOWS system make instrument operation more convenient.
14. Electromagnetic interference is shielded and can work normally even near mobile phones or two-way wireless communication devices.
DeltaClassicDE-2000 alloy analyzer element analysisrange
All alloys ranging from titanium (Ti) to plutonium (Pu) can be analyzed.
| Analysis Mode and Element Types of Innov-X Delta DE-2000 (Classic Type) Alloy Analysis Instrument | |
| Analysis | Analysis Elements |
| Scope of Elemental Analysis | 31 basic elements ranging from element 22 titanium to element 94 PU, within which other elements can be replaced according to customer needs. |
| Alloy mode | Ti,V,Cr,Mn,Fe,Co,Ni,Cu,Zn,Hf,Ta,W,Re,Pb,Bi,Zr,Nb,Mo,Pd,Ag,Cd,Sn,Sb Waiting for 23 elements |
| project | (Classic) Alloy Analyzer Innov-X Delta DE-2000 |
| Size and weight | Dimensions: 245x250x88mm; Weight<1.5KG |
| Environmental requirements | Environmental humidity ranging from 0% to 95%; Environmental working temperature -20 ℃~50 ℃; Blue gray, black, silver white alternating |
| excitation source | High power micro straight plate electronic X-ray tube with an excitation voltage of 35kV; no high-voltage cables, no RF noise, better X-ray shielding, and better heat dissipation. |
| X-ray tube target material | Au/Ta target |
| x-ray detector | High sensitivity Si Pin detector |
| Cooling system | Adopting a Peltier constant temperature cooling system, the controller operates at -35 ℃ to ensure the detection accuracy of the instrument and is not affected by external temperature |
| filter | Fixed filter |
| Voltage and current | Voltage 35kV, current 100uA |
| Host power supply system | 2 lithium batteries, 110v/220 universal charger, charger adapter, intelligent docking station |
| Intelligent docking station | Additional batteries can be charged, and the built-in battery of the instrument can be charged simultaneously and display the charging progress. The docking station can be connected to a computer to exchange data, allowing the instrument to be standardized in real time and on standby at any time |
| Power on and replace the battery | The instrument can replace the battery without shutting down even when turned on, with a time of up to 90 seconds |
| standardization | The instrument does not require standardization when turned on and can be directly tested. Standardization is only an optional option |
| balance | The instrument has good balance and can stand on the workbench during testing without the need for manual support. With a one click button design, it can operate for a long time without fatigue |
| Heat dissipation | More than one-third of the machine body adopts an aluminum alloy shell design, and there is a dedicated slot type heat dissipation device on the top of the instrument. The entire system makes heat dissipation very effective, prolongs machine life, and the X-ray analyzer works more stably, resulting in an extremely low failure rate |
| monitor | Integrated design of the whole machine, industrial grade high-resolution TFT QVGA Casio BlanView ® Touch display color screen with LED backlight; Pixel size 240 × 320 |
| Display fixing method | All in one machine design, integrated structure of the whole machine, non removable PDA, dustproof, anti fog, waterproof, low fault |
| Display visibility | The monitor has no LCD altitude sickness and can have excellent visibility even in low light and strong light environments indoors. It has low energy consumption and is half as low as traditional white light displays, increasing display performance in various environments |
| data shows | The percentage (%) displays the element content, and the element display order can be sorted by energy, concentration value, user-defined, etc. It can also calculate the average value of multiple tests and can be connected to a desktop computer for display; The instrument synchronously and dynamically displays the chemical composition during the testing process |
| Data Storage | ROM 128M, RAM 128M, 2G SD card, capable of storing 205000 sets of data and spectra |
| data transmission | USB cable and wireless Bluetooth for data transfer, files can be output in TXT and EXCEL formats. |
| Processor CPU | 532MHz CPU, Freescale, ARM1136-MX31 processor, floating-point operation mode, significantly improved speed |
| System CPU peripherals | Adopting USB bus, I2C, GPIO, Bluetooth, accelerator, real-time clock chip, micro SD memory card (capable of storing 2GB data), GPIO strip connected to trigger and PSM |
| operating system | User oriented Windows CE 6.0 system |
| Alloy mode | Ti,V,Cr,Mn,Fe,Co,Ni,Cu,Zn,Hf,Ta,W,Re,Pb,Bi,Zr,Nb,Mo,Pd,Ag,Cd,Sn,Sb Waiting for 23 elements |
| Alloy modeling method | Basic parameter method, the instrument does not require prior knowledge of the alloy type before analyzing the alloy, achieving fully automated analysis |
| Type of alloy | Ferroalloy series, nickel based alloy series, cobalt based alloy series, titanium based alloy series, copper based series, 6 High temperature alloy, molybdenum tungsten alloy, hybrid alloy series |
| Alloy grade | There are over 500 standard alloy grades in the standard library, with up to 10000 types of alloys that can be analyzed. Users can customize 300 alloy grades |
| Small sample testing | Compensatory testing methods for irregular or small samples can detect very small or few samples, such as filaments with a diameter of 0.04mm that can be immediately identified |
| hot test | High temperature materials capable of detecting temperatures up to 450 ℃ |
| Comparison of actual sample test values for 304 | |||||
| Reading | Mode | Cr | Mn | Fe | Ni |
| Standard value | Analytical | 18.56 | 1.53 | 71.24 | 8.09 |
| Actual measurement value | Analytical | 18.66 | 1.57 | 71.16 | 7.85 |
| Actual measurement value | Analytical | 18.61 | 1.58 | 71.02 | 8.05 |
| Actual measurement value | Analytical | 18.62 | 1.52 | 71.18 | 7.98 |
| Actual measurement value | Analytical | 18.68 | 1.57 | 71.12 | 7.91 |
| Actual measurement value | Analytical | 18.67 | 1.5 | 71.03 | 7.99 |
| Actual measurement value | Analytical | 18.62 | 1.55 | 71.22 | 7.93 |
| Actual measurement value | Analytical | 18.5 | 1.57 | 71.2 | 7.96 |
| Actual measurement value | Analytical | 18.7 | 1.53 | 71.14 | 7.99 |
| Actual measurement value | Analytical | 18.63 | 1.52 | 71.31 | 7.88 |
| Actual measurement value | Analytical | 18.65 | 1.51 | 71.23 | 7.95 |