Online conductivity electrodeIt simplifies the functions while ensuring performance, thus having a particularly strong price advantage. Clear display, easy operation, and excellent testing performance make it highly cost-effective. It can be widely used for continuous monitoring of conductivity values in solutions such as thermal power, chemical fertilizers, metallurgy, environmental protection, pharmaceuticals, biochemistry, food, and tap water.
Performance Characteristics
Microcomputerization: Using high-performance CPU chips, high-precision AD conversion technology, and SMT surface mount technology to complete multi parameter measurement, temperature compensation, and automatic range conversion, with high accuracy and good repeatability.
High reliability: touch sensitive buttons, no switch knob or potentiometer.
Automatic conversion of measurement frequency: avoids electrode polarization and improves measurement accuracy.
Phase sensitive detection design: eliminate the influence of wires on measurement.
25 ℃ conversion: temperature compensation automatic measurement/manual input.
Alarm function: Alarm signal isolation output, alarm upper and lower limits can be set arbitrarily, alarm hysteresis cancellation.
Industrial control watchdog: ensures that the instrument does not crash.
Network function: isolated current output and RS485 communication interface.
Isolation current output: The upper and lower limits of the conductivity value corresponding to the current can be set arbitrarily.
Adopt flow type, sinking type, flange type or pipeline installation.
Online conductivity electrodeTechnical Specifications
| Conductivity measurement range | 0.01-20 μ S/cm (with K=0.01 electrode); 0.1-200 μ S/cm (with K=0.1 electrode); 1.0-2000 μ S/cm (with K=1.0 electrode); 10-20000 μ S/cm (with K=10.0 electrode); |
| Basic error of electronic unit | Conductivity ± 0.5% FS, temperature ± 0.3 ℃; |
| Automatic temperature compensation range | 0-99.9 ℃, with 25 ℃ as the reference temperature; |
| Basic instrument error | Conductivity ± 1.0% FS, temperature ± 0.5 ℃ |
| Repetitive error of electronic unit | ± 0.2% FS ± 1 word; |
| Stability of electronic unit: | ± 0.2% ± 1 word FS/24h; |
| Electronic unit alarm error | ≤±3.0% FS; |
| Error caused by the influence of environmental temperature on electronic units | ≤±0.5% FS; |
| Error caused by the influence of electronic unit power supply voltage | ≤±0.3% FS; |
| Current isolation output | 4-20 mA (load<750 Ω); |
| Electronic unit output current error | ≤±0.5% FS; |
| Alarm relay | AC220V 、3A; |
| RS485 communication interface | (Optional); |
| power supply | AC220V ± 22V, 50Hz ± 1Hz; or 24V |
| Protection level | IP65; |
| Overall dimensions | 96 (length) x 96 (width) x 110 (depth) mm, |
| cutout size | 92×92 mm; |
| weight | 1.0kg ; |
| working conditions | Environmental temperature 0-60 ℃, relative humidity<85%; |
| Flow pool inlet and outlet connection pipeline | Hard or flexible pipes, with outer diameters available in three specifications: diameter 8, diameter 10, and diameter 12; |
| Can be equipped with four conductivity electrode constants | 0.01, 0.1, 1.0, 10.0. |
When measuring the conductivity electrode constant, the following three points should be noted:
1. When measuring, it should be done.
2. The temperature of the KCL solution for measuring electrode constants should be close to the actual temperature of the measured solution.
3. The concentration of KCL solution for measuring electrode constants should be close to the actual concentration of the measured solution.