The online dust particle counter OPC-6303UL adopts the principle of optical scattering, which can accurately detect and calculate the number of suspended particles of different sizes in the air per unit volume. It is equipped with a four-way photoelectric dust source intelligent recognition module and a flow stable air pump, which can simultaneously output the number of particles in six channels: 0.3 μ m, 0.5 μ m, 1.0 μ m, 2.5 μ m, 5.0 μ m, and 10 μ m (default unit: pcs/m); , switchable pcs/L, pcs/ft3), with indication error meeting JJF 1190 standard.
Online dust particle counterOPC-6303UL adopts the principle of optical scattering, which can accurately detect and calculate the number of suspended particles of different sizes in the air per unit volume. It is equipped with a four sided photoelectric dust source intelligent recognition module and a flow stable air pump. It can simultaneously output the number of particles in six channels: 0.3 μ m, 0.5 μ m, 1.0 μ m, 2.5 μ m, 5.0 μ m, and 10 μ m (default unit: pcs/m ³, switchable between pcs/L and pcs/ft3), and the indication error meets the JJF 1190 standard.
Online dust particle counterOPC-6303ULProduct Features
Intelligent particle recognition algorithm to meet accurate measurements in different atmospheric environments
Industrial grade laser with high reliability
Constant current sampling structure ensures a constant sampling flow rate
Can output particle counts of 0.3 μ m, 0.5 μ m, 1.0 μ m, 2.5 μ m, 5.0 μ m, and 10 μ m (default pcs/m ³, pcs/L, pcs/ft3)
| Detection Principle |
Principle of Light Scattering |
| number of channels |
六通道 (0.3μm、0.5μm、 1.0μm、2.5μm、5.0μm、10μm) |
| sampling flow |
2.83L/min |
| Counting efficiency |
50%@0.3μm 100% @ ≥ 0.5 μ m (25 ± 2 ℃, 50 ± 10% RH environmental conditions) |
| Power on stability time |
≤8s |
| Data refresh rate |
1s |
| working conditions |
0 ℃~45 ℃: 0~95% RH (non condensing) |
| Storage conditions |
-20~60 ℃, 0~95% RH (non condensing) |
| Working Voltage |
DC 12-24V |
| Average working current |
≤1A |
| communication interface |
RS485 interface (standard) RJ45 (standard) 4-20mA (standard) |
| display |
3.5-inch color screen |
| Alarm light (optional) |
Warning method: LED flashing light Tone: siren sound Sound level: 110 decibels Protection level: IP44 Working mode: 485 signal control alarm |
| Temperature and humidity accuracy |
Temperature: ± 1 ℃ Humidity: ± 8% RH |
*Please refer to the specification sheet for specific parameters. If you need more technical information, please contact us!