E-mail
2259938953@qq.com
-
Phone
18656860316
-
Address
Room 608, Building 1, No. 348 Huajiang Road, Jiading District
Shanghai Fatai Precision Instrument Co., Ltd
2259938953@qq.com
18656860316
Room 608, Building 1, No. 348 Huajiang Road, Jiading District
FT-E22typeOxygen hydrogen analysis detectorUsing imported sensors with sensitive response and high accuracy, combined with a signal processing circuit designed with a low-power chip, the oxygen content in the measured gas is converted into4-20mA standard signal output, and digital display of the hydrogen content of the measured gas; The operation adopts touch buttons, which are simple and reliable; Adopting a two-wire system and 24V DC power supply; Small size, light weight, reasonable structure, and easy installation; Suitable for online detection and analysis of hydrogen content in hydrogen and other inert gases in various explosion-proof environments. For example, the process of electrolyzing water to produce hydrogen.
FT-E22typeOxygen hydrogen analysis detectorThe transmitter complies withThe relevant provisions of GB3836.1-2000 and GB3836.4-2000 standards have been passed the inspection by the National Explosion proof Electrical Product Quality Supervision and Inspection Center; Explosion proof mark: Exia Ⅱ CT6
applicationField:
Online detection and analysis of hydrogen content in industrial production processes such as fertilizers, petrochemicals, electrolysis of water for hydrogen production, and biological fermentation for explosion-proof and non explosion proof applications.
Instrument features:
At the bottomImported electrochemical sensors have the characteristics of long lifespan, high anti-interference ability, high precision, and fast response
At the bottomOn site LCD display, easy to operate for maintenance and calibration, ensuring long-term measurement accuracy
At the bottomSmall size, light weight, easy to install
At the bottomHigh reliability intrinsic safety explosion-proof design,IP65 protection level
At the bottomOne time table adopts standards4-20mA two-wire signal output, long transmission distance
Following standards:
At the bottomJJG 663-1990 Verification Regulations for Thermal Conductivity Hydrogen Analyzer
Technical Specifications:
At the bottomMeasuring range:0.00~2.00% H2 (customizable)
At the bottomessence Degree:≤±1%F. S
At the bottomheavy again Sex:≤±0.5%
At the bottomdivide distinguish Rate:0.01%
At the bottomResponse time:T90 ≤ 10 seconds
At the bottomSample gas pressure:0.05 MPa ≤ inlet pressure ≤ 0.1 MPa (outlet must be at atmospheric pressure)
At the bottomSample gas flow rate:400±10 mL/ min
At the bottomOutput signal:4-20mA (two-wire system)
At the bottomWorking power supply:DC24V
At the bottomWorking environment: Operating temperature: -25℃~+55℃
At the bottomOperating humidity:≤ 90% RH (no condensation)
At the bottomDimensions: Transmitter:140mm (width) x 140mm (height) x 80mm (depth)
At the bottomExplosion proof mark:ExiaⅡCT6
At the bottomheavy Quantity: Approximately10kg
At the bottomExternal safety barrier parameters:Vo≤28VDC、Io≤93mA、Cp≤0.02uF、Lp≤1mH
Principle Description
Electrochemical gas analyzer is a chemical gas analysis instrument. It measures gas composition based on changes in ion concentration or current caused by chemical reactions. In order to improve selectivity, prevent surface contamination of measuring electrodes, and maintain electrolyte performance, a diaphragm structure is generally used. The commonly used electrochemical analyzers include positioning electrolysis and galvanic cell.