The all-new ELEMENTRAC OH-p 2 is a powerful elemental analyzer used to measure oxygen and hydrogen concentrations in inorganic samples such as steel, iron, copper, or ceramics. High sensitivity non dispersive infrared detectors and thermal conductivity detectors can detect element contents ranging from very low ppm to high percentage, providing stable and reliable results. The innovative injection system features pulse degassing and vertical injection design, making it easy to analyze rod-shaped, granular, and powdered samples with analysis weights up to 2g. The ELEMENTRAC OH-p2 elemental analyzer can meet or even exceed the requirements of all relevant international standards, such as ASTM E 1019 or DIN EN 3976.
Alloy, aluminum, dust, carbides, cast iron, copper, ferroalloy, iron, metal, ore, refractory metal, silicon, steel
Step 1: Enter the sample information into the Elements software
The sample name is entered into the software, and the weight information will be automatically transferred to the software (see step 2)
ELEMENTRAC OH-p 2 can accurately analyze sample sizes ranging from a few milligrams to 2 grams. Rod shaped or granular samples can be directly analyzed. For elemental analysis of powders, nickel capsules do not need to be sealed for analysis.
ELEMENTRAC OH-p 2 can be used as a single element analyzer to analyze oxygen or hydrogen separately, or to measure both oxygen and hydrogen simultaneously. Among them, oxygen is measured in the form of carbon dioxide in two infrared detectors, while nitrogen and hydrogen are measured in elemental form in a thermal conductivity cell.
The chemical reagents and filters required for the operation of the ELEMENTRAC OH-p 2 element analyzer are conveniently located at the front of the instrument panel and can be hidden behind a movable door during daily operation. This arrangement greatly reduces the time required for maintenance and increases the overall experience. In addition, innovative details greatly improve the reproducibility of measurements.
Innovative sample inlet&pulse degassing sample chamber
The new sample inlet of the OH-p2 element analyzer ensures convenient operability and reproducibility. Samples of different shapes, such as solid blocks, particles, or powders in capsules, can be analyzed with a weight of up to 2000 milligrams, and the surrounding air is removed through high-pressure degassing at the injection port. Then they are vertically placed into the preheated graphite crucible for analysis.
Closed gas management
The ELEMENTRAC ONH elemental analyzer series uses a closed gas system under high pressure to ensure that 100% of the released sample gas is sent to the detector, ensuring good sensitivity and reproducibility.
In addition to the one-stop solution of ELEMENTRAC OH-p 2 elemental analyzer, it also provides further options to improve analysis efficiency and expand application scope.
Automatic cleaning device
Melting the sample in a graphite crucible at temperatures up to 3000 ° C can cause carbon deposition in the upper electrode and furnace, which will have a negative impact on the repeatability of ONH testing. The new optional automatic cleaning instrument can reliably remove these carbon deposits and achieve accurate elemental analysis even under high-throughput conditions. In addition, the elemental analyzer also has efficient gas calibration function and carrier gas pre cleaning function in furnace cleaning.
The comprehensive element analysis software based on Windows is an essential component of all element generation element analyzers. The middle window (analysis and results) is the starting point for all the functions required for daily work. From here, analyzed samples can be grouped and exported, or new samples can be registered and analyzed. Users can call various dependent functions, such as application settings, calibration, diagnosis, or status.
| Analysis Elements | Oxygen, hydrogen |
| sample | inorganic |
| Furnace calibration | vertical |
| Sample loading | Graphite crucible |
| application area | Steel, metallurgy, ceramics, metals, engineering, electronics |
| stove | Pulse furnace power up to 8.5 kW, temperature up to 3000 ℃ |
| detection | Infrared detection method for measuring oxygen; Thermal conductivity method for measuring hydrogen |
| Typical analysis time | 120 – 180 s |
| chemicals | Schuetze reagent, sodium hydroxide, magnesium perchlorate |
| carrier gas | Compressed air, nitrogen purity 99.995%, all gases (2-4 bar/30-60 psi) |
| Power requirements | 3~ 400 V, 50/60 Hz, Power 8500 W |
| Equipment dimensions (width x height x depth) | 56 x 78 x 64 cm |
| weight | ~ 165 kg |
| Required equipment | Computer, monitor, balance (accuracy 0.0001 grams) |
| Select accessories | Cooling cycle unit, gas standard unit, carrier gas purification furnace |
| – | *6.8 kW during application setup |