Brief introduction: The oxygen rich spark ignition test device complies with the requirements of clause 11.2.2 of 9706.1-2020 and clause 11.2.2.1 of IEC60601-1:2015,
Can be used for spark ignition testing of medical equipment in oxygen rich environments. Oxygen rich environments refer to situations where the ambient pressure does not exceed 110 kPa,
Concentration greater than 25%; Or when the environmental pressure is greater than 110kPa, the oxygen partial pressure is greater than 27.5kPa

Equipment parameters
(1) 1 copper electrode with a diameter of 3mm, 1 copper electrode with a diameter of 1mm
(2) Adjustable DC power supply, voltage adjustable from DC0-100V
(3) Resistance adjustable from 0-200 ohms
(4) Inductance 0-12mL adjustable
(5) Capacitor 0-1500uF adjustable
(6) Digital voltmeter 0-500V, accuracy 0.5%, display accuracy 0.1V
(7) Digital ammeter 0-5A, accuracy 0.5%, display accuracy 0.01A
(8) Power factor table PF 0-1, display accuracy 0.01
(9) Probe movement speed 0-20mm/s, speed accuracy 0.1m/s
(10) Spark ignition frequency 1-9999 can be set
(11) The ignition time can be set from 0-9.99s
(12) Oxygen flow meter 0-10L/min 2% FS
(13) The human-machine interface adopts Siemens touch screen, and the control adopts Siemens PLC
5. Equipment principle: The oxygen flow rate is controlled by a flow meter, and the probe is moved by a stepper motor to drive the lead screw,
Accurately ensure full contact of the probe. The outer frame is made of organic glass for easy observation of experimental phenomena.
Equipment parameters
(1) 1 copper electrode with a diameter of 3mm, 1 copper electrode with a diameter of 1mm
(2) Adjustable DC power supply, voltage adjustable from DC0-100V
(3) Resistance adjustable from 0-200 ohms
(4) Inductance 0-12mL adjustable
(5) Capacitor 0-1500uF adjustable
(6) Digital voltmeter 0-500V, accuracy 0.5%, display accuracy 0.1V
(7) Digital ammeter 0-5A, accuracy 0.5%, display accuracy 0.01A
(8) Power factor table PF 0-1, display accuracy 0.01
(9) Probe movement speed 0-20mm/s, speed accuracy 0.1m/s
(10) Spark ignition frequency 1-9999 can be set
(11) The ignition time can be set from 0-9.99s
(12) Oxygen flow meter 0-10L/min 2% FS
(13) The human-machine interface adopts Siemens touch screen, and the control adopts Siemens PLC
5. Equipment principle: The oxygen flow rate is controlled by a flow meter, and the probe is moved by a stepper motor to drive the lead screw,