Ozone aging resistance test chamberStructural characteristics:
1Inner wall: importedSUS304stainless steel
2Shell: Cold rolled plate electrostatic spraying
3Sealing: Imported magnetic sealing strip
4Temperature and humidity control method:BTHCBalance temperature and humidity regulation
5The sample rack consists of an aluminum frame type lining plate and a tensile plate
6Korean intelligent digital display controller, ozone concentration automatic digital display controller. PIDSet adjustment,0.2Grade accuracy,LEDDisplay, with communication interface
7Heating: Stainless steel fin air heating
8Ozone generator: High voltage silent discharge tube
9Variable speed motor driven sample holder device
(Satisfy the static tensile test,The running trajectory of the specimen conforms to the principle of cycloid trajectory)
10Convection system: Aluminum fan blades for electric motors(定制)
11The device runs a timer and automatically shuts down when it ends
12Double beam;QUA;Non dispersive UV ozone detector, Bill-Lambert measurement application software.
13Static strain test method,
14Dynamic strain test method.

Safety protection device for ozone aging test chamber:
1 Power overload, control circuit overload, short circuit protection
2 Grounding protection, motor overload protection
3 Over temperature protection
4 Overvoltage and overload protection
5 To protect the equipment,All alarms will automatically cut off the power and sound a warning.
On site conditions for using the ozone aging test chamber:
1Temperature:15℃~35℃
2Relative humidity: not greater than85%RH
3There is no strong vibration or electromagnetic field around
4There is no high concentration of dust or corrosive substances around
5Without direct sunlight or direct radiation from other heat sources
6There is no strong airflow around, and when the surrounding air needs to be forced to flow, the airflow should not be directly blown onto the box.
7The test chamber should be placed steadily and kept level.
8There should be a certain distance around the test chamber to facilitate maintenance operations.
9The installation site has good ventilation
10Good grounding
The main technical indicators of the ozone aging test chamber are:
1.Temperature range:RT+10~65℃
2.Temperature uniformity: ±2℃
3.Temperature fluctuation: ±0.5℃
4.Ozone concentration:10PPHm~1000PPHm
5.Ozone concentration deviation: ±10%
6.Material rack rotation speed20~25mm /s
7.Sample tensile frequency:0.5±0.25Hz (This indicator is only applicable to dynamic ozone aging test chambers)
8.New sample holder(Each fixture can be used to install test pieces, and the tensile strength can meet the requirements of Type I dumbbell pads and actual workpieces)
9.Test gas flow rate:0-60L/min
Ozone aging resistance test chamberOur services:
1Provide technical information such as electrical schematics, consumable parts lists, operation and maintenance manuals for the equipment, as well as technical information for the main purchased supporting components.
2Train equipment operators and maintenance personnel to master equipment operation skills and general repair and maintenance skills.
3From the date of acceptance of the equipment1Provide free after-sales service to buyers within the year
4. (Long term free support for software and technology upgrades) Paid services will be provided after the warranty period expires.
The device is manufactured based on one or a combination of the following standards:
GB 7762-2003Ozone aging resistance test for vulcanized rubber Static tensile test method
GB/T 13642-92Ozone aging resistance test for vulcanized rubber Dynamic tensile testing method
ISO 1431/2-1982vulcanized rubber-Resistant to ozone cracking-Part Two:Dynamic strain test
Aisili InstrumentIt is early production in ZhuchengHigh temperature sterilization potThe enterprise!Developed byHigh temperature sterilization potBecome a creative product in the industry!Produced byHigh temperature sterilization potperformance,The offline qualification rate is as high as!Complete after-sales system!