Matters required before testing the YZS-10A fully automatic oxygen index tester——Xinsheng Zhuorui YZS-10A Oxygen Index Tester has a high degree of automation,The automatic oxygen index tester of Beijing Xinsheng Zhuorui Technology Co., Ltd. is applicable to the combustion performance measurement of wire and cable, plastic, rubber, fiber, foam plastic, film and film, textile and other materials. It is a new type of product developed according to the technical requirements specified in GB/T2406. The fully automatic oxygen index analyzer is used to measure the volume percentage of oxygen required during the combustion process of polymers. The polymer oxygen index value is the low oxygen concentration (also known as oxygen index) required to maintain combustion for 50mm or 180s in the oxygen and nitrogen mixture airflow after the substance is ignited. The fully automatic oxygen index analyzer can not only be used as a means of identifying the flame retardancy of polymers, but also as a research tool, providing powerful testing methods for laboratory research on flame retardant formulas and the development of new flame retardant materials.
Precautions to be taken before testing the fully automatic oxygen index analyzerDesign standard for fully automatic oxygen index analyzer: GB/T 2406.2-2009 GB/T 2406.1-2008
The fully automatic oxygen index tester meets the standards of ASTM D 2863, ISO 4589-2, NES 714 GB/T 5454 GB/T 10707 GB/T 8924-2005 GB/T 16581-1996 NB/SH/T 0815-2010 TB/T 2919-1998 IEC 61144-1992 ISO 15705-2002 ISO 4589-2-1996

YZS-10A fully automatic oxygen index tester with touch screen control (with smoke exhaust hood) Product advantages:
1. Full color touch screen control, simply set the oxygen concentration value on the touch screen, and the program will automatically adjust to the oxygen concentration balance and emit a beep sound prompt, eliminating the trouble of manually adjusting the oxygen concentration; One click calibration of oxygen concentration and data storage;
2. The step proportional valve is used to greatly improve the flow control accuracy. The closed-loop control is used to automatically adjust the oxygen concentration drift program to the target value during the test, which avoids the disadvantage that the traditional oxygen index tester cannot adjust the oxygen concentration during the test, and greatly improves the test accuracy;
3. The equipment comes with a fume hood that can automatically exhaust smoke;
YZS-10A Fully Automatic Oxygen Index Tester Touch Screen Control (with Smoke Exhaust Hood) Main Technical Parameters:
1. Adopting imported oxygen sensors, the digital display of oxygen concentration does not require calculation, is accurate, and has a range of 0 to
2. Display accuracy ± 0.01
3. Touch screen setting program automatically adjusts oxygen concentration
4. One click calibration accuracy
5. One click concentration ratio
6. Automatic alarm sound for stable oxygen concentration
7. Equipped with timing function
8. Can store experimental data
9. Historical data can be queried
10. Can clear historical data
11. Can choose whether to burn 50mm or not
12. Gas source fault prompt
13. Oxygen sensor fault prompt
14. Oxygen and nitrogen misconnection prompt
15. Aging reminder for oxygen sensor
16. Standard oxygen concentration input
17. The diameter of the combustion cylinder can be set (two commonly used specifications are optional)
18. Flow regulation range: 0-20L/min (0-1200L/h)
19. Quartz glass tube: Choose one of two specifications (inner diameter ≥ 75mm or inner diameter ≥ 85mm)
20. Gas flow velocity inside the combustion cylinder: 40mm ± 2mm/s
21. Overall dimensions of the machine: 80cm * 40 * 102cm
22. Outer diameter of automatic smoke exhaust duct: 7.6cm
23. Pressure gauge accuracy level 2.5, resolution: 0.01MPa
24. Experimental environment: ambient temperature: room temperature to 40 ℃; Relative humidity: ≤ 70%;
25. Input pressure: 0.25-0.3MPa
26. Working pressure: Nitrogen 0.15-0.20Mpa, Oxygen 0.15-0.20Mpa
27. Propane (butane) ignition system, the ignition nozzle is made of a metal tube, and the tail end has a nozzle with an inner diameter of Φ 2 ± 1mm, which can be freely bent. Can be inserted into the combustion cylinder to ignite the sample, flame length: 16 ± 4mm, size 5mm-60mm can be freely adjusted,
28. Gas: Industrial nitrogen, oxygen, purity>99%; (The purity of industrial oxygen provided by users/regular gas stations is generally between 99.2% and 99.3%).
29. Power requirements: AC220 (+10%) V, 50HZ
30. High power consumption: 150W
31. Self supporting material sample clamp: can be fixed at the axis position of the combustion cylinder and can vertically clamp the sample
32. Non self-supporting material sample clamp: capable of fixing both vertical edges of the sample to the frame at the same time (optional/applied to soft non self-supporting materials such as textiles)

33. Upgradeable to IoT functionality: Control and read data through a router, save data to a computer, and print reports Taken from: Beijing Xinsheng Zhuorui Technology Co., LtdZhuorui YZS-10A Fully Automatic Oxygen Index Tester
Precautions to be taken before testing the fully automatic oxygen index analyzer