E-mail
qiyatest@163.com
-
Phone
13537151533
-
Address
Baishigang Third Industrial Zone, Changping Town, Dongguan City
Dongguan Qiya Testing Equipment Co., Ltd
qiyatest@163.com
13537151533
Baishigang Third Industrial Zone, Changping Town, Dongguan City
Technical Scheme for Thermal Aging Test Chamber
*Control methods and characteristics:
Balanced temperature and humidity control system(BTHC), in order toP.I.D.mode controlSSRMake the heating capacity of the system equal to the moisture and heat loss, so it can be used stably for a long time.
heat aging test chamberUsed for conducting thermal aging tests on plastic materials (rubber, plastic), electrical insulation, and other materials in scientific research institutions and factories. conform toGB/T 3512-2014.ISO188; two thousand and elevenWaiting for national standards.
1、 Performance:At room temperature of 20 ℃, when unloaded
1.1Temperature range:RT+10℃~300℃
1.2Temperature fluctuation:±0.5℃
1.3 Temperature uniformity:±2℃
1. 4 Air exchange rate: 2-100 times/H
1.5 Timing devices0~999hour
1.6 Wind speed: adjustable from 0m/s to 2.0m/s
1.7test stand:rotational speed10RPM/minTest rack
2、 Structure and Material:
machine type |
within part chi cun |
outside part chi cun |
QYD-150L |
500x600x500(WxHxD)MM |
850x1600x900(WxHxD)MM |
2.1Inner box material:Stainless steel plate
2.2External material:SPCC board with electrostatic baking paint
2.3Insulation material:glass wool
2.4Supply air circulation system:
2.4.1 60WMotor,1only
2.4.2Stainless steel extended axis
2.4.3Multi wing fan blade(SIRCCO FAN)
2.4.4Adjustable louvers ensure even temperature distribution inside the test chamber
2.5Box door:Single panel door, left opening, handle on the right-hand side
2.5.1Flat embedded handle
2.6ventilation system:
2.6.1 25WMulti wing snail fan
2.6.2A code tuner
2.6.3One adjustable air door
3、 Control system:
3.1 temperature controller:Original digital display button controller
3.2 power meter:Digital display button power meter
3.2 Heating system:
3.2.1heater:Nickel chromium alloy finned heat dissipation tubeUHigh efficiency heater with automatic calculation and high precision
linear measurePID+SSRControl (calculation cycle:1~4S can be set up)Make its heat loss equal to the heating amount
Comparable, therefore more than traditionalON/OFFControl power saving approximately30%
3.3 Temperature measuring body: DINspecificationslevel ψ4.8mm SUS #304Made of stainless steelPT 100Ω 1branch
3.4 Safety device:
3.4.1Zero crossing gate fluid power controller
3.4.2Air burning prevention protector
3.4.3Underphase reverse phase protection switch
3.4.4 Line fuses and fully sheathed terminals
3.4.5 buzzer
4、 Standard configuration:
4.1 power line4.0 mm²x5Rui, low impedance rubber cable2.5mlong1strip
4.2 One set of controller user manual and one set of operation manual
5、 Surrounding environment:
5.1Allowable temperature range for use0~35℃
5.2Performance guarantee scope:5~35℃
5.3 Relative humidity: not greater than85%
5.4Atmospheric pressure:86~106Kpa
5.5There is no strong vibration around
5.6No direct sunlight or other heat sources
6、 Use power supply:AC 3 µ 4W380V 50HZ; Power: 6.0KW