1、 Product Introduction of Ammonia Nitrogen Water Quality Online Automatic Monitoring Instrument
1Overview of online automatic monitoring instrument for ammonia nitrogen water quality
The ammonia nitrogen water quality online automatic monitoring instrument is currently mainly used in surface water, sewage, tap water, rainwater, rivers, farmland, enterprise production, environmental inspection and other places. The instrument can automatically complete data monitoring of on-site water sources, and transmit data to various required places such as mobile phones, computers, and the internet. The instrument has the functions of accurate and fast results, environmental protection, automatic continuous monitoring, and data networking.
2. Method basis for online automatic monitoring of ammonia nitrogen water quality
Water quality - Determination of ammonia nitrogen - Salicylic acid spectrophotometric methodExecution standardsHJ 101-2019
ammonia nitrogen Technical requirements for automatic water quality monitoring instrumentHJ/T 101-2003
Brief Introduction to the Principle of Ammonia Nitrogen Water Quality Online Automatic Monitoring Instrument:
After mixing the water sample and masking agent, the ammonia nitrogen formed as free ammonia or ammonium ions reacts with salicylic acid colorimetric reagent in an alkaline environment and in the presence of sensitizers to generate a colored complex. The analyzer detects the color change and converts it into ammonia nitrogen value output. The amount of colored complex generated is equivalent to the ammonia content.
3. Typical application of online automatic monitoring instrument for ammonia nitrogen water quality
Design requirements for monitoring station buildings of polluting enterprises:
Five parameters of water quality (temperature, pH, dissolved oxygen, conductivity, turbidity)
Online analyzer for COD/ammonia nitrogen/total phosphorus/total nitrogen/permanganate/heavy metals, etc
Data acquisition instrument, industrial computer, dynamic control (uploaded to * * * data center)
Water sample collection system, customized water distribution system, fully automatic water quality sampler system, rainfall and dust sampling system
Monitoring station building and supporting facilities, fire protection facilities, lightning protection facilities and other all-weather monitoring systems
2、 Product parameters of online automatic monitoring instrument for ammonia nitrogen water quality
Analyze the range |
(0-2) mg/L, (0-10) mg/L, (0-50) mg/L (expandable range) |
|
detection limit |
0.02mg/L |
|
resolution |
<0.01mg/L |
|
Error |
20% FS± 8.0% 50% FS± 5.0% 80% FS± 3.0%
|
|
repetitiveness |
≤5% |
|
Span Drift |
±1.0%F.S.
|
|
dissolution temperature |
165 ℃ (adjustable) |
|
Duration of resolution |
15 minutes (adjustable) |
|
analysis method |
The whole point |
Running by the hour, 24 sets per day |
Interval |
Run at set intervals |
|
automatic |
automatic analysis |
|
long-range |
Remote control through MODBUS mode; Dry contact control |
|
Calibration method |
interval count |
Perform interval calibration based on the cumulative number of analyses |
days interval |
Calibrate the interval based on the number of running days |
|
Manual/Automatic |
Manual calibration/automatic calibration |
|
long-range |
Remote control via MODBUS |
|
Cleaning Method |
automatic |
Analysis and cleaning; Fault cleaning; Abnormal power outage cleaning; Emergency stop cleaning |
hand manual |
manual cleaning |
|
call the police |
Liquid shortage alarm; Self checking alarm; Fault alarm; Over limit alarm (one relay) |
|
data |
A 4-20mA analog signal |
|
Communication |
1 RS232 digital interface or RS485, supporting MODBUS communication protocol or custom protocol |
|
Storage |
About 20000 sets (automatically saved upon power failure); Support exporting EXCEL from USB drive; |
|
Micro industrial thermal printer (optional) |
||
Maintenance Cycle |
About 30 days, it takes 45 minutes |
|
display unit |
Touch screen (7 inches) 800 * 480 resolution, TFT true color touch screen |
|
Usage environment |
5—35℃, Humidity (20~85)% RH (No condensation) |
|
Water sample temperature |
5—35℃ |
|
Water collection unit |
Used for the circulation and retention of water samples; Automatic Backwash Precision Filtration System (Optional) |
|
Size |
420mmx420mm x1370mm (L x W x H)
|
|
weight |
45KG (including reagent cabinet) |
|
power supply |
AC 220V ± 10%, 50Hz ± 1% |
|
power |
150W |
|
3、 Product Features of Ammonia Nitrogen Water Quality Online Automatic Monitoring Instrument
1. The main components of the instrument are imported to ensure its accuracy, stability, and reliability;
2. Infrared unit, accurate quantification, capable of completing multiple quantitative calculations;
3. Adopting positive/negative pressure current carrying technology, the peristaltic pump tube does not pass through liquid and is stable and durable.
4. Three point automatic calibration with high linear correlation;
5. Dedicated LED light source, high signal-to-noise ratio;
6. RS232/RS485 interface, standard MODBUS protocol, can remotely read the operating status and related data of the instrument;
4、 Ammonia nitrogen water quality online automatic monitoring instrument service
Instrument purchase process; Notify requirements, confirm on-site environment, provide instrument scheme design, sign contract, deliver goods, install and construct, provide on-site training, provide reagent configuration methods (customers can purchase and configure by themselves), transfer materials, and complete networking procedures
The warranty period for the entire instrument is 12 months. If there are any quality problems with the instrument during the warranty period, free replacement, repair, and technical support will be provided.
The manufacturer provides maintenance services for the products sold;
1. Consumables are sourced from * * standard parts, with reasonable and transparent prices;
2. Formula disclosure, provision of reagent and standard solution preparation services, reagents using AR analytical purity, * * * drug quality;
3. Exclusive application solutions, on-site optimized design;
4. After sales technical support;