1Product Overview of RM-8000A Ammonia Nitrogen Online Analyzer
RM-8000A Ammonia Nitrogen Online AnalyzeryesforSurface water andemissionsewageThe ammonia nitrogen indexOnline monitoring instruments.instrumentwith***StandardorBased on industry standards,haveEasy to operate, highly automated, and easy to maintainConsumablesLow cost characteristics,instrumentStandard configuration includesRS485/RS232Interface, supportsMODBUSProtocol, it is possiblepassremotecommunicationRegarding its operational statusandComprehensive data analysismaster,Can alsoRemote control of it to achieve remote analysis、remotecalibrationRemote cleaning, etc**Function.in additioninstrumentEquipped with dilution function, it can also meet the application requirements for * * * samplesofrequirement. sutraprolongedFeedback verification,Its analysisThe degree of * *, etcindicatorSuperior * * relatedStandard. This product is suitable for:Surface water, domestic sewage, industrial wastewater, etc.
IIRM-8000A Ammonia Nitrogen Online AnalyzerProduct Features
1.The main components are imported to ensure their accuracy, stability, and reliability;
2.Infrared unit, precise quantification;
3.Adopting positive/Negative pressure current carrying technology, peristaltic pump tube does not pass through liquid, stable and durable.
4.Three point automatic calibration with high linear correlation;
5.dedicatedLED light source, high signal-to-noise ratio;
6.RS232/RS485 interface, standard MODBUS protocol, can remotely read the operating status and related data of the instrument;
IIIRM-8000A Ammonia Nitrogen Online AnalyzerTechnical Specifications
Measurement method:Distillation Nessler spectrophotometric colorimetric method HJ 535-2009 | |||
Test range:(04)mg/l,(015)mg/l,(050)mg/l,(0300)mg/lfourRange of gearsAuto Switch | |||
The testing range after automatic dilution can be extended to:01500mg/l | |||
Offline detection:0.05mg/l | |||
Resolution: <0.02mg/l | |||
Accuracy:Standard solution<10%; water sample<15% | |||
Reproducibility: < 5% | |||
resolveTime:20min, can be set | |||
Fault free running time:≧720h/time | |||
Range drift:±5%F. S. | |||
sampleInterval:Continuous1 hour, 2 hours... 24 hours, trigger, * * time point | |||
Correction interval:Manually or automatically at selected intervals and times (1-7 days) | |||
Cleaning interval:Manually or automatically at selected intervals and times (1-7 days) | |||
Maintenance interval:>1Every month, everytimeabout1hour | |||
Reagent consumption:Each set of reagents is approximately720About one sample | |||
human-machine interface:7inch70000 colors, 800 * 480 resolution, TFT true colortouch screen | |||
Printing:Reserved printer interface for external industrial useMini printer (optional) | |||
Storage: 2Ten thousand pieces of data, no loss in case of power failure, automatically overwrite early data when fully stored(Can add 40000 pieces of data) | |||
|
Communication interface: 1 RS232 digital interface or RS485, supporting MODBUS communication protocol or custom protocol 1 analog signal 4-20mA (20mA corresponds to adjustable range) | |||
Preprocessing system: self-cleaning, blowback, precision filtration functionWhile ensuring good representativeness of the samples, it also avoidslarge-scaleSuspended particles block the pipelineEnsure the continuity of data.(optional) | |||
Overall dimensions |
900×600× 450(mm) |
weight |
50kg |
power supply |
AC 220V ± 20%, 50Hz ± 1% |
power |
500W |
ambient temperature |
5~40℃ |
ambient humidity |
≤85% |