The insulation material voltage breakdown tester is mainly suitable for testing the breakdown strength and withstand time of solid insulation materials such as plastics, films, resins, mica, ceramics, glass, insulation oil, insulation paint, cardboard and other media under power frequency voltage or DC voltage; The voltage breakdown tester is controlled by a computer, and the test curve can be observed online during the test process; Automatically store test conditions and test results data, and can access, display, and print them
Product Name:Voltage breakdown tester
Product Model: SDJ-10KV SDJ-30KV SDJ-50KV SDJ-100KV SDJ-150KV
Insulation material voltage breakdown testerMeet the standards:
1. GB1408.1-2006 "Test Methods for Electrical Strength of Insulation Materials"
2. GB1408.2-2006 "Test Methods for Electrical Strength of Insulation Materials
Part 2: Additional Requirements for Application of DC Voltage Test
3. GB/T1695-2005 "Determination method for power frequency breakdown voltage strength and withstand voltage of vulcanized rubber"
4. GB/T3333 Test Method for Power Frequency Voltage Breakdown of Cable Paper
5. HG/T 3330 Determination of breakdown strength of insulation paint film
6. GB12656 Test method for paper frequency voltage breakdown of capacitors
7. ASTM D149 "Test Method for Dielectric Breakdown Voltage and Dielectric Strength of Solid Electrical Insulation Materials at Industrial Power Supply Frequency"
Insulation material voltage breakdown testerMainly applicable to solid insulation materials such as:Testing of breakdown strength and withstand time of media such as plastics, films, resins, mica, ceramics, glass, insulating oil, insulating paint, cardboard, etc. under power frequency voltage or DC voltage;
voltage breakdownThe testing instrument is computer-controlled, and the test curve can be observed online during the testing process; Automatically store test conditions and test results data, and can access, display, and print them
Technical Indicators:
1. Input voltage:AC220V/50Hz
2. Output voltage:AC:0~50kV; DC:0~50kV
3. Output power:5kVA
4. Measurement range:AC:0~50kV; DC:0~50kV
5. Measurement error: ≤ 2%
6. Boosting rate: 1kV/s to 5kV/s
7. Voltage endurance time: 0-4H
8. The leakage current of 1-30 mA can be freely set by computer software
9. Power supply AC 220V ± 10% single-phase AC voltage and 50Hz ± 1% frequency
10. Test environment temperature: 15-30 ℃, relative humidity: 0-85%, can operate stably.
11. Dimensions: length x width x height: 1000mm×750 mm×1350mm
12. Equipment weight:150Kg (reference)
13. Grounding requirements: The instrument needs to be separately grounded, and the grounding should comply with national standards. The metal rod should be buried at least 1.5 meters deep underground