E-mail
ryu@ultrablue-sci.com;info@ultrbalue-sci.com
-
Phone
13331917708
-
Address
Room 1019, Building 3, Lane 1000, Lingshan Road, Pudong New Area, Shanghai
Shanghai Beilan Optoelectronics Technology Co., Ltd
ryu@ultrablue-sci.com;info@ultrbalue-sci.com
13331917708
Room 1019, Building 3, Lane 1000, Lingshan Road, Pudong New Area, Shanghai
The optical system of Spectis1.0 adopts a miniature collimating lens with a working wavelength range of 340-780nm. It is a miniature low stray light system that provides a data acquisition interval of 1.7nm, making it very suitable for measuring the performance of light sources such as LED and OLED.
Spectis1.0Featuring low noise and high-speed data transmission, the built-in temperature sensor monitors changes in ambient temperature to ensure the stability of measurement data. Software calibration programs can reduce the effects of stray light, noise, etc., ensuring measurement accuracy.

Applicable to:
Lighting product manufacturer
Lighting Design Engineering Company
Lighting testing agency
Environment or personnel related to lighting inspection and testing
System indicators
Lux/Lux - irradiance
Lumen/Lumen - Luminous flux
CRI - Color rendering index (compliant with CIE standards)
CCT - Color Temperature (in accordance with the standard CIE)
Color - Chromatic Coordinates (CIE 1931 and CIE 1964)
Fidelity and Color Gamut - Method for Evaluating Light Source Color Restoration According to TM-30 IES Standard
PAR/PPFD - photosynthetically active radiation measurement (horticulture)
EML - Equivalent Black Luminous Lux/Equivalent Black Luminous Illumination
MWatt - Radiant Energy
systemspecificationsparameter
Spectral Range* |
340-780nm / 340-740nm LS / 640-1050nm |
||||
detector |
CMOS Sensor |
||||
pixel |
256 |
||||
physical resolution |
~ 1.7nm |
||||
Wavelength reproducibility |
±0.5nm |
||||
integration time |
10ms~10s (in automatic mode) |
||||
A/D conversion |
16bits |
||||
Signal to Noise Ratio |
1000:1 |
||||
stray |
2*10E-3 |
||||
Spectral radiation accuracy** |
6% (in the range of 200-220 nm), 5% (in the range of 220-500 nm), 4% (in the range of 500-1050 nm) |
||||
Measurement uncertainty of color coordinates (x.y)** |
0.0015 |
||||
Illuminance (lux) |
10 lx~20000 lx (for white LED) |
||||
Luminous flux [lm] |
✓(Together with the scoring ball) |
||||
Brightness [cd/m ²] |
✓useGL OPTI PROBE detector together |
||||
USB |
USB2.0 |
||||
operation temperature |
5-35°C |
||||
Dimensions [H x W x D] |
115mm x 72mm x 19mm |
||||
weight |
120g |
||||
*The spectral range of the sensor may be reduced due to the limitations of the optical accessories used, which may lower the actual spectral range of the system. |
|||||
**Absolute measurement uncertainty immediately appears after calibration. The expanded uncertainty corresponds to a coverage probability of 95%, with a coverage factor of k=2. | |||||
|
This parameter is effective under laboratory conditions of 25 ° C and relative humidity of 45%.
be careful:Instrument, firmware, and software specifications are subject to change without prior notice.All information contained in the GL OPTIC data sheet, as well as any form of product information, has been carefully prepared and includes truthful information. Please note that differences may occur due to text and/or other errors or changes in available technology. It is recommended to contact us before using the product to obtain Z's new product specifications. | |||||