low-speed centrifugesummary:
TDL-40Blow-speed centrifugeStable performance, large capacity, low noise during operation, small temperature rise inside the centrifugal chamber, and beautiful appearance. Separation technology is a technique developed based on the state of particles in a practical centrifugal environment. This machine adopts a DC brushless motor, controlled by a microcontroller, and can pre select speed, time, centrifugal force, digital display, timed shutdown sound prompt and fault sound prompt, with multiple protections such as lock and cover opening shutdown. The button panel design is easy to operate, with fast start and stop, and automatic balancing function. Widely applicable to laboratories in universities, petroleum, chemical, pharmaceutical, environmental protection and other industries
Product features:
·Adopting all steel structure and stainless steel centrifugal chamber.
·Microcomputer control, DC brushless variable frequency motor drive, smooth and quiet operation.
·Touch panel and digital display.
·Parameters can be changed at any time during operation without stopping the machine.
·Automatically calculate and set the centrifugal force RCF value, with the RCF key available for observation at any time.
·When the countdown time is less than one minute, it is displayed in seconds.
·Using mechanical door locks, reliable and durable.
·The fault self diagnosis system detects faults such as overspeed and speed loss during operation, and immediately decelerates and stops the machine to improve the safety performance of the instrument.
·Applicable: Suitable for biochemical laboratories, hospital clinical testing, etc
TDL-40Bdesktop
Technical Specifications:
Rotational speed |
5000rpm(turn/divide) |
relative centrifugal force |
4390(×g) |
Horizontal rotation子capacity |
100ml×4 (5000rpm) |
50ml×8 (5000rpm) | |
|
10mlor15ml×16(5000rpm) 10mlor15ml×32(5000rpm) | |
timing range |
0min~99min |
power supply |
220v 50Hz 8A |
Overall dimensions |
510mm×410mm×310mm(L×W×H) |
weight |
30kg |
Main technical parameters:
1. Power supply: AC 220V 50HZ
2. Overall power: 600W
3. Speed: adjustable at 5000 revolutions per minute
4. Centrifugal capacity: 4 × 100ml, 8 × 50ml, 16 × 10/15mlor32×10/15ml
5. Relative centrifugal force: 4390xg
6. Timing range: 1-99 minutes
Usage:
1. Pressright sideOpen the door cover with the lock button, install the rotor and test tube, and ensure that the testing capacity in the centrifuge tube is not greater thantotal90% of the capacity is symmetrically placed in the rotor body, the door cover is closed, and the power is turned on, Turn on the power switch.
2. Add the centrifuge tube equipped with this machine to the substance to be separatedqualityInsert them into the centrifuge head holes separately, and determine the amount of substance added to each centrifuge tubeheavyThe quantity must be basically equal. Close the cover plate, press the selection button, and the rotor number window will flash again. Use ▲▼ to select the rotor number, press the selection button again, and the speed window will flash. Use ▲▼ to select the speed you need, then press the selection button again, and the working time window will flash again. Use ▲▼ to select the time you need, press the (approve) button to confirm, and then press the start button. At this time, the centrifuge enters the working state. It will automatically emit when the set working time is reachedTimed shutdown sound prompt.
3. Press the start switch, and the whole machine will start running. If you need to stop halfway, please press the stop button. The machine will automatically stop when it reaches the set time. At this time, the buzzer will sound. After the rotor stops steadily, you can open the back cover, take out the sample, and turn off the power switch after use.
maintenance and upkeep:
1. The drive shaft should be coated with a small amount of butter, Vaseline, or lubricant to prevent rust and facilitate the removal of the rotor.
2. If the centrifuge tube ruptures or centrifugal liquid drops into the rotor, it must be washed and dried with clean water in a timely manner.
3. Centrifugal tubes should be promptly updatedchange.
4. After use, the water stains in the centrifuge chamber should be wiped dry, and the dirt in the centrifuge chamber and on the rotor should be cleaned. The rotor should be taken out and placed in a clean and dry place.
5. When the centrifuge is not in use for a long time or during maintenance, the main power plug should be unplugged, otherwise the instrument will be charged. During maintenancecanAn accident occurred.
6. It is strictly prohibited for any liquids or other debris to enter the centrifuge chamber. otherwisecanDamaged the host.
7. The instrument casing should be properly grounded to prevent accidents caused by moisture damage to the entire machine.
8. Please do not open the upper cover of the centrifuge head at high speed, as it will automatically stop when opened.
9. When there is an error in selecting recognition, it will automatically occurSound alertRemind you to reselect to confirm health