Welcome Customer !

Membership

Help

Hangzhou Tiankang Instrument Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Products

Hangzhou Tiankang Instrument Co., Ltd

  • E-mail

    tkyb2005@126.com

  • Phone

    18072820595

  • Address

    No. 536 Shunfeng Road, Qianjiang Economic Development Zone, Yuhang District, Hangzhou City

Contact Now

TK-3051TG pressure transmitter

NegotiableUpdate on 10/27
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

TK-3051TG pressure transmitter material diaphragm: stainless steel 316L, Hastelloy C process connection: stainless steel 316L filling liquid: silicone oil transmitter housing: aluminum alloy material, coated with epoxy resin outer shell sealing ring: nitrile rubber (NBR) nameplate: stainless steel 304 .....

Product Details

The CHD820 series gauge/absolute pressure transmitter is used to measure the liquid level, density, and pressure of liquids, gases, or vapors, and then convert them into 4-20mADC HART current signal output. It can also communicate with HART375 or HART475 handheld controllers for parameter setting, process monitoring, etc.
standard specifications
(Calibration range based on standard zero point, stainless steel 316L diaphragm, filled with silicone oil)
1: Performance specifications
Reference accuracy for range adjustment(including linearity, hysteresis, and repeatability starting from zero)±0.075%
If TF>10 (TF=maximum range/adjustable range), then it is:±(0.0075×TF)%
Optional maximum accuracy of 0.05%.
Environmental temperature impact

Range code

-Total impact from 20 ℃ to 70 ℃

B

±(0.20×TF+0.10)%×Span

other

±(0.10×TF+0.10)%×Span

long-term stability

Range code

impact quantity

B

± 0.2% × Span/1 year

other

± 0.1% × Span/1 year

Power supply impact
± 0.001%/10V (15-42V DC), negligible
2: Functional specifications
Range and Range (CHD820-TG gauge pressure)

Range/Range

KPa

bar

B

range

0.6~6

6~60mbar

scope

-6~6

-60~60mbar

C

range

2~40

0.02~0.4

scope

-40~40

-0.4~0.4

D

range

2.5~250

0.025~2.5

scope

-100~250

-1~2.5

F

range

30~3000

0.3~30

scope

-100~3000

-1~30

G

range

0.1~10MPa

1~100

scope

-0.1~10MPa

-1~100

H

range

0.20~20 MPa

2.0~200

scope

-0.1~20 MPa

-1~200

I

range

0.4~40 MPa

4~400

scope

-0.1~40 MPa

-1~400

J

range

0.6~60 MPa

6~600

scope

-0.1~60 MPa

-1~600

Range and Range (CHD820-TA Absolute Pressure)

Range/Range

KPa

bar

B

range

2~40

0.02~0.4

scope

0~40

0~0.4

C

range

2.5~250

0.025~2.5

scope

0~250

0~2.5

E

range

30~3000

0.3~30

scope

0~3000

0~30

Range limit
Within the upper and lower limits of the measuring range, it can be adjusted arbitrarily. Suggest selecting range codes with the lowest possible range ratio to optimize performance characteristics.
Zero point setting
The zero point and range can be adjusted to any value within the measurement range in the table, as long as the calibration range is ≥ the minimum range
Installation location impact
Changes in the installation position parallel to the membrane surface will not cause zero drift effects. If the installation position changes by more than 90 degrees from the membrane surface, there will be zero drift within the range of<0.25kPa for range C, and zero drift within the range of<0.15kPa for other ranges. Both can be corrected by adjusting the zero position. No range influence.
output
2-wire system, 4-20mADC, optional HART output digital communication, optional linear or square root output.
Output signal limit: Imin=3.8mA, Imax=20.8mA
alarm current
Low alarm mode (minimum): 3.8mA
High alarm mode (maximum): 20.8 mA
Non reporting mode (hold): Maintain the effective current value before the fault
Alarm current standard setting: high alarm mode
response time
The damping constant of the amplifier component is 0.1s; the time constant of the sensor is 0.1-1.6s, depending on the range and range ratio. The additional adjustable time constant is 0.1 to 60 seconds.
Preheating time:<15s
ambient temperature
-40~85℃
-20~70 ℃ with LCD display and fluororubber sealing ring
Storage temperature/transportation temperature
-50~85℃
When equipped with LCD display: -40~85 ℃
Pressure limit
From vacuum to maximum range.
Overload limit:

range

6kPa (B)

40kPa (C)

250kPa (D)

3MPa (F)

Overload limit

200KPa

500kPa

2000KPa

7.5MPa

range

10MPa (G)

20MPa (H)

40MPa (I)

60MPa (J)

Overload limit

20MPa

40MPa

60MPa

70MPa

Electromagnetic Compatibility (EMC)
Refer to the "Electromagnetic Compatibility Appendix" on the next page
3: Installation
Power supply and load conditions
The power supply voltage is 24V, R ≤ (Us-12V)/Imax k Ω
Where Imax=23 mA
Maximum power supply voltage: 42VDC
Minimum power supply voltage: 12VDC, 15VDC (backlit LCD display)
Digital communication load range: 230-600 Ω
electrical connection
M20X1.5 cable sealing buckle, the terminal block is suitable for wires of 0.5-2.5mm2.
process connection
Standard process connection: NPT 1/2 internal thread, can be converted to NPT 1/2, G1/2, M20x1.5 external thread, KF16 vacuum interface.
4: Physical specifications
material
Diaphragm: Stainless steel 316L, Hastelloy C
Process connection: Stainless steel 316L
Filling fluid: Silicone oil
Transmitter housing: made of aluminum alloy material, coated with epoxy resin on the outside
Shell sealing ring: Nitrile rubber (NBR)
Nameplate: Stainless Steel 304
weight
1.6kg (none: LCD display, mounting bracket, process connection)
Shell protection levelIP67
Electromagnetic Compatibility Appendix

serial number

test project

Basic standards

test conditions

performance level

1

Radiation interference (shell)

GB/T 9254-2008 Table 5

30MHz~1000MHz

qualified

2

Conducted interference (DC power port)

Table 1 of GB/T 9254-2008

0.15MHz~30MHz

qualified

3

Electrostatic discharge (ESD) immunity

GB/T 17626.2-2006

4kV (contact)
8kV (air)

B

4

Radio frequency electromagnetic field immunity

GB/T 17626.3-2006

10V/m (80MHz~1GHz)

A

5

Power frequency magnetic field immunity

GB/T 17626.8-2006

30A/m

A

6

Electrical fast transient burst immunity

GB/T 17626.4-2008

2kV(5/50ns,5kHz)

B

7

Surge immunity

GB/T 17626.5-2008

1kV (between lines)
2kV (between line and ground) (1.2us/50us)

B

8

Immunity to Conducted Interference Induced by Radio Frequency Field

GB/T 17626.6-2008

3V(150KHz~80MHz)

A

Note: (1) A Performance Level Description: During testing, the performance is normal within the technical specification limits.
(2) B Performance Level Description: During testing, the functionality or performance may temporarily decrease or be lost, but it can recover on its own without changing the actual operating conditions, storage, and data.

Dimensions Unit (mm)
Horizontal piping connection method (side) Horizontal piping connection method (front)
TIANKANG
Wall mounted connection methodVertical piping connection method
TIANKANG
5: Electrical connection diagram
TIANKANG
6: Process Connection Instructions (Code 5)
6.1 Standard Form (Code N)
TIANKANG
6.2 Derivative Interface Forms
1/2 inch NPT external thread (code A) M20x1.5 external thread (code M)
TIANKANG
G 1/2 external thread (code G) vacuum interface DIN 28403 KF16/ISO 2861 (code Y)
TIANKANG
1/4 inch NPT external thread (code Y)
TIANKANG
7: CHD820 series intelligent pressure and absolute pressure transmitter model and specification code
CHD820-TG Intelligent Pressure Transmitter Selection Spectrum

CHD820-TG

CHD820-TG Intelligent Pressure Transmitter

Code 1

range

overvoltage

B

0-0.6KPa~6KPa

200KPa

C

0-2KPa to 40KPa

500kPa

D

0-2.5KPa to 250KPa

2000KPa

E

0-10KPa to 1000KPa

3.0MPa

F

0-30KPa to 3000KPa

7.5MPa

G

0-0.1MPa~10MPa

20MPa

H

0-0.2MPa~20MPa

40MPa

I

0-0.4MPa~40MPa

60MPa

J

0-0.6MPa~60MPa

70MPa

Code 2

output

E

Standard intelligent (4~20) mA DC two-wire system+HART protocol digital communication

Code 3

diaphragm

1

316L stainless steel

2

Hastelloy C

Code 4

Fill liquid

2

silicone oil

3

fluorinated oil

Code 5

Process interface

N

1/2NPT internal thread

A

1/2NPT external thread

G

G1/2 external thread

M

M20 × 1.5 external thread

Y

Special connection method

Code 6

electrical interface

N

1/2NPT internal thread

M

M20 × 1.5 internal thread

Code 7

mounting bracket

B4

Pipe support (carbon steel)

B5

Flat bracket (carbon steel)

B6

Pipe support (stainless steel)

B7

Flat bracket (stainless steel)

Code 8

Explosion-proof rating

N

Ordinary type (without explosion-proof)

D

Explosion proof Exd Ⅱ CT6

I

Intrinsic safety explosion-proof Exia Ⅱ CT6

Code 9

accuracy class

2

Level 0.2

5

Level 0.5

1

0.1 level

7

0.075 level

6

0.065 level

8

0.05 level

Option code

option

K

Degreasing treatment

L

Hanging license plate

F

Lightning protection (withstand transient voltage)

H

Equipped with ¢ 14 welded pressure pipe (stainless steel)

E

English nameplate

CHD820-TA series intelligent absolute pressure transmitter selection chart

CHD820-TA

CHD820-TA Intelligent Absolute Pressure Transmitter

Code 1

range

overvoltage

C

0-2~40KPa

500kPa

D

0-2.5~250KPa

2000KPa

E

0-10~1000KPa

3.0MPa

F

0-30~3000KPa

7.5MPa

Code 2

output

E

Standard intelligent (4~20) mA DC two-wire system+HART protocol digital communication

Code 3

diaphragm

1

316L stainless steel

2

Hastelloy C

Code 4

Fill liquid

2

silicone oil

3

fluorinated oil

Code 5

Process interface

N

1/2NPT internal thread

A

1/2NPT external thread

G

G1/2 external thread

M

M20 × 1.5 external thread

Y

Special connection method

Code 6

electrical interface

N

1/2NPT internal thread

M

M20 × 1.5 internal thread

Code 7

mounting bracket

B4

Pipe support (carbon steel)

B5

Flat bracket (carbon steel)

B6

Pipe support (stainless steel)

B7

Flat bracket (stainless steel)

Code 8

Explosion-proof rating

N

Ordinary type (without explosion-proof)

D

Explosion proof Exd Ⅱ CT6

I

Intrinsic safety explosion-proof Exia Ⅱ CT6

Code 9

accuracy class

2

Level 0.2

5

Level 0.5

1

0.1 level

7

0.075 level

6

0.065 level

8

0.05 level

Option code

option

K

Degreasing treatment

L

Hanging license plate

F

Lightning protection (withstand transient voltage)

H

Equipped with ¢ 14 welded pressure pipe (stainless steel)

E

English nameplate