This instrument complies with the national standard GB/T 11142-89 and the national industry standard NB/SH/T 0810-2010 "Determination of Gas Evolution of Insulating Oil under Electric Field and Ionization". After the insulating oil is dried and saturated with hydrogen gas, the insulating liquid and the hydrogen layer on the liquid surface are subjected to the action of a radial electric field under the conditions of voltage of 10KV, frequency of 50Hz, oil temperature of 80 ℃, and testing time of 120min. The interface between oil and hydrogen gas tends to absorb or release gas due to discharge reflection.
TP575 Insulation Oil Gassing Tester Oil Gassing * Tester
lWidely used in petrochemical, power, railway, scientific research and other departments, it is an indispensable equipment for oil analysis and quality inspection.
l7-inch large touchscreen LCD screen, with clear images and easy operation.
lCombining different standards, customers have high selectivity.
lBuilt in flowchart for easy user experimentation and operation.
lImported temperature sensor with high measurement accuracy.
lImported temperature control module, high-precision temperature control.
lEquipped with an oil drain valve, the oil change operation is convenient.
lThe thermal printer produces stable and reliable printing results.
lEquipped with safety protection switches, the instrument is safe and reliable to use.
lStore 1000 pieces of historical data for easy retrieval.
TP575 Insulation Oil Gassing Tester Oil Gassing * Tester
Relative humidity:<85%
Heating power: ≤ 1500W
Mixing speed: 1200 revolutions per minute
Power supply voltage: AC 220V ± 10% 50Hz ± 10%
Whole machine power: ≤ 1700W
External dimensions:
Controller: 320mm × 305mm × 195mm
High voltage generator: 320mm × 305mm × 380mm
Gas evolution tester: 320mm × 305mm × 590mm
Hydrogen inlet pressure: 0.05-0.1Mpa
Constant temperature: 80 ℃± 0.05 ℃ (adjustable range: room temperature~
100℃)
Resolution: 0.01 ℃
Test voltage: 10kV ± 0.2 kV
Resolution: 0.01kV
Time measurement: 5 minutes, 10 minutes, 50 minutes, 120 minutes
(Automatically converted according to standards)
Timing error: < ± 0.1s
Operating temperature: (10-40) ℃
lAccessories Guide
lGas separation pool
lMeasuring tube
Precautions
1. The instrument casing should have good contact with the ground to ensure safety.
When there is no liquid in the constant temperature bath, do not start the instrument, otherwise it will be damaged
Bad heater.
When replacing fuses or other components, the power plug should be unplugged.
If a new measuring tube is replaced, data needs to be re entered.
5. Non professionals are not allowed to dismantle or repair instruments at will.
After using the instrument, the power should be cut off in a timely manner.