Instrument Overview:
The fully automatic closed flash point tester is a high-precision automated equipment used to determine the flash point value of liquid fuels such as petroleum products and lubricants under closed conditions1It is based onGB/T 261-2021Designed according to standards, it quickly locks the flash point value and generates reports through intelligent temperature control, automatic ignition, and other methods. It is widely used in fields such as power, petroleum, chemical, inspection, and scientific research.
This instrument adopts microcontroller control technology, color LCD display screen, fully Chinese human-machine dialogue interface, and unmarked keyboard; It has menu prompts and guided input functions for parameters such as predictable flash point temperature, sample label, atmospheric pressure, and test date; Apply open, fuzzy control integrated software, modular structure design and other technologies. This instrument is designed with advanced technology, user-friendly interface, easy operation, and accurate test results. It is an ideal substitute for similar imported instruments and can be widely used in railway, aviation, power, petroleum industries, as well as in colleges, research institutes, metrology and testing departments for detecting and testing the flash point and ignition point of petroleum products.
Technical parameter accuracy:
project |
scope |
Operation interface: |
10Inch industrial control computer, switching between Chinese and English |
control system |
PLC+windows |
Test cup |
comply withGBT261-2021 |
Automatic ignition system |
Caliber:0.7-0.8mm |
heating rate |
1-6℃/min(Adjustable) |
stirring rate |
30r/min-250r/min |
skateboard |
Thickness approximately2.4mm |
Measurement method: |
Thermocouple differential detection formula |
Measuring range: |
room temperature-400℃ (To determine the flash point at minus zero temperature, the sample needs to be frozen first) |
divide distinguish Rate: |
0.1℃ |
barometer |
range30-110KPa,Accuracy: Shi0.5Kpa |
Repeatability: |
National standard: flash point value<104Allowable deviation at ℃ ±1℃, flash point value ≥104Allowable deviation at ℃ ±2℃ |
printer: |
Low energy high-speed thermal printer,56Millimeter paper width |
Temperature detection: |
Imported platinum resistor(Pt100Ω) measure |
temperature accuracy |
110Below ℃ is0.5℃,110Above ℃ is1℃ |
Ignition method: |
Fully automatic electronic ignition. |
optional |
Heating bath or oven |
Whole machine power consumption: |
not greater than500W. |
Overall dimensions |
long570 wide400tall732 mm |
Instrument Features
1This instrument uses colorDisplay, fully Chinese human-machine dialogue interface, unmarked keyboard, with prompt menu and guided input function for parameters such as pre value temperature, sample label, atmospheric pressure, and test date.
2Simulate and track the function curve of temperature rise and test time, with Chinese misoperation software prompt modification function, equipped with parameter prompt functions such as test date and test time.
3Automatically calibrate the influence of atmospheric pressure on the experiment and calculate the correction value.
4Differential detection, automatic correction of system deviations.
5The scanning, detection, and printing of data are automatically completed, and the test arm is automatically raised and lowered.