The touch screen salt spray testing machine adopts an automatic or manual water addition system, which can automatically replenish the water level when the water level is insufficient, and the test is not interrupted. The precise glass nozzle (PYREX) disperses the mist evenly and naturally falls on the test piece, ensuring no crystal salt blockage. Equipped with dual over temperature protection and low water level warning to ensure safe use.
For the touch screen salt spray testing machine, look for Dongguan Aisili Testing Instrument. Zhong S, 20 years old, has strong technology and meets all the standards of salt spray testing machines. Aisili salt spray test chamber, salt spray corrosion resistance testing machine, salt spray machine is trustworthy!
Touch screen salt spray testing machineFeatures:
Adopting an automatic or manual water addition system, it can automatically replenish the water level when the water level is insufficient, and the test will not be interrupted. The precise glass nozzle (PYREX) disperses the mist evenly and naturally falls on the test piece, ensuring no crystal salt blockage. Equipped with dual over temperature protection and low water level warning to ensure safe use.
The temperature controller uses digital display and PID control, with an error of ± 0.1 ℃.
The laboratory adopts steam direct heating method, which heats up quickly and evenly, reducing standby time.
The conical disperser of spray tower has the functions of guiding fog, adjusting fog volume and even falling fog volume.
The saturated air tank adopts the Henry's law to heat and humidify, and provides the required humidity for the laboratory.
Salt water concentration: 5%; The pH value of water is between 6.5-7.2.
Touch screen salt spray testing machineTechnical Specifications:
Model SH-60 SH-90 SH-120 SH-160
Weight 60kg 80kg 120kg 160kg
Machine external dimensions (W × H × D) cm 110 × 60 × 118 150 × 90 × 130 190 × 130 × 140 230 × 130 × 140
Laboratory temperature NSS ACSS 35 ℃± 1 ℃ (CASS 50 ℃± 1 ℃)
Saturated air tank temperature NSS ACSS 47 ℃± 1 ℃ (CASS 63 ℃± 1 ℃)
Laboratory space (W x D x H) cm 60 x 45 x 40 90 x 60 x 50 120 x 100 x 50 160 x 100 x 50
Power supply AC220V 1 Φ 20A AC220V 1 Φ 20A AC220V 1 Φ 30A AC380V 3 Φ 15A
Laboratory volume 110L 270L 600L 800L
Saltwater volume 15L, 25L, 40L, 65L
Material
The laboratory uses PVC polyvinyl chloride board
The laboratory seal adopts acrylic impact resistant board with a thickness of 5mm
The reagent replenishment bucket adopts a hidden water level observation window, which is easy to clean and not easy to break
The saturated air tank adopts SUS # 304 stainless steel high-pressure resistant insulation effect
The blue frame in the laboratory adopts a flat dividing frame, which can adjust the angle freely. In terms of fog and fog resistance in all corners, there are many test pieces placed