Welcome Customer !

Membership

Help

Beijing Puxi General Instrument Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Products

Beijing Puxi General Instrument Co., Ltd

  • E-mail

    yafei.wei@pgeneral.com.cn

  • Phone

    13681197768

  • Address

    No. 3 Pingsan Road, Pinggu District, Beijing

Contact Now

Universal TAS-986 Single Beam Atomic Absorption Spectrometer

NegotiableUpdate on 09/19
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The universal TAS-986 single beam atomic absorption spectrometer, as an important means of determining trace metal elements, has the advantages of high sensitivity, good accuracy, and fast analysis speed. It has been widely applied in various fields such as geology, metallurgy, materials, medicine, environment, and chemistry, and is now becoming a very important instrument analysis method.

Product Details

As a professional manufacturer of analytical instruments, Beijing Puxi General Company has been committed to the research and development of atomic absorption spectrometers for many years. In 1997, it launched the highly automated TAS-986 atomic absorption spectrophotometer in China, which uses fully Chinese Windows operating software. All controls of the instrument are completed by computers, greatly reducing the difficulty of operation for analysts, avoiding human errors, and ensuring high analytical accuracy. Received a gold award at the BCEIA Awards in 1997.


Universal TAS-986 Single Beam Atomic Absorption SpectrometerKey Features:
1. The structural design of an integrated flame atomizer and graphite furnace atomizer has achieved automatic switching between flames and graphite furnace atomizers
2. Domestic companies adopt advanced horizontal heating graphite furnace technology to ensure uniform temperature of graphite tubes, reducing chemical interference and memory effects
3. Adopting a rotating lamp frame with eight lamps, automatically selecting the position of each element's hollow cathode lamp
4. Automatically set * * height position and gas flow rate to ensure * * analysis conditions; Automatic wavelength scanning, peak searching, and spectral bandwidth replacement
5. The flame monitor can monitor flame changes at any time. In case of accidental flameout, the instrument will automatically shut down the acetylene flow and prompt an alarm

Technical Specifications:
1. Wavelength range: 190nm-900nm
2. Wavelength accuracy: ± 0.25nm
3. Wavelength repeatability: 0.15nm
4. Heating temperature range: 20 ℃~2650 ℃ (horizontal heating, which is equivalent to 3000 ℃ for vertical heating)
5. Spectral bandwidth: automatic switching between five levels: 0.1, 0.2, 0.4, 1.0, 2.0
6. Resolution: better than 0.3nm
7. Baseline drift: 0.005A/30min

spectroscopic system

wavelength range 190nm-900nm
monochromator C-T type monochromator device for eliminating aberration
spectral bandwidth 0.1, 0.2, 0.4, 1.0, 2.0nm: five automatic switching modes
wavelength accuracy ±0.25nm
Wavelength repeatability accuracy 0.15nm
resolution Better than 0.3nm
baseline drift 0.005A/30min

Performance specifications of universal TAS-986 atomic absorption spectrophotometer for general analysis

Characteristic concentration (Cu) 0.03µg/ml/1%
Detection limit (Cu) 0.006µg/ml
burner Metal titanium burner
precision RSD≤1%
sprayer Efficient glass atomizer
雾化室 Corrosion resistant material atomization chamber
Position adjustment Automatic setting of flame burner height and front and rear positions
safety measures Equipped with multiple automatic protection functions

Graphite furnace analysis

Characteristic quantity (Cd) 0.5×10-12g
Detection limit (Cd) 1×10-12g
precision RSD≤3%
Heating temperature range Room temperature~2650 ℃
Heating and temperature control method The power control method for the drying and ashing stage and the high power control method for the atomization stage are controlled by light
Heating condition setting 9 programs for ramp heating, step heating, and high-power heating

Background correction

Deuterium lamp background correction Correctable 1A background
Self suction background correction Correctable 1A background

data processing

measurement method Flame method, graphite furnace method, hydride generation atomic absorption method
Concentration calculation method Standard curve method (1-3 degree curve), standard addition method; Interpolation method
Repeated measurement times Calculate the average of 1-20 times and provide the standard deviation and relative standard deviation
Result printing Parameter printing, data result printing, graphic printing

Universal TAS-986 Single Beam Atomic Absorption SpectrometerCore parameters

Optical system: Single beam Detector: PMT
Monochrome components: echelle Background deduction: comprehensive
Repeatability: ≤0.15nm Detection limit: 0.006 μ g/mL (flame mode); 1.0×10-12g