The existing vacuum source (such as a basic pump or a powerful vacuum network) combined with the VACUU · SELECT desktop vacuum controller can improve experimental efficiency. VACUU · SELECT vacuum controller can meet all your requirements. For simple experimental processes, control parameters can be manually modified, pre-defined applications can be launched with just one click, or one can create their own applications through simple "drag and drop" operations. This series of controllers can be used as desktop devices, installed on laboratory racks, or embedded in laboratory vacuum networks. Integrated VACUU · SELECTdesktop-styleThe vacuum controller includes a VACUU · SELECT control panel, one-way check valve, and chemically resistant electromagnetic isolation valve. The ceramic vacuum sensor and vent valve are directly installed on the VACUU · SELECT vacuum controller, with a compact configuration that is easy to install. This vacuum controller features an intuitive, application based user interface that provides pre-defined vacuum processes for all common laboratory applications. During the solvent evaporation process, the controller can detect the pressure when the solvent boils and use a two-point control method to maintain the pressure in a stable state.
◆ Fully configured and ready to use, including integrated valves and universal hose connectors
Pre defined applications and automatic boiling point detection function can save time
◆ Simple setup - Connect the VACUU · SELECT vacuum controller between the pump and the application
The integrated vent valve allows the connection of inert gas
The integrated one-way check valve reduces the mutual influence between the same vacuum network and other applications
| technical indicators | unit | VACUU·SELECT |
|---|---|---|
| Vacuum sensor | VACUU·SELECT Sensor | |
| Upper limit of measurement range | mbar/torr | 1080 / 810 |
| Lower limit of measurement range | mbar/torr | 0.1 / 0.1 |
| Measuring principle | Ceramic diaphragm (alumina), capacitive, independent of gas type and pressure | |
| Measurement uncertainty | <+-1 mbar/hPa/torr/+-1 digit (calibrated, constant temperature) | |
| temperature coefficient | < 0.15 mbar/hPa/0.11 torr /K | |
| Air release valve, interface | Integrated, DN 4-5 mm hose nozzle connector | |
| Environmental temperature range (operation) | °C | 10 - 40 |
| Environmental temperature range (storage) | °C | -10 - 60 |
| Medium temperature for continuous operation/short-term operation | °C | 40 / 80 |
| Shell material | A sturdy plastic casing with excellent chemical corrosion resistance | |
| Protection level | IP 40 | |
| The protection level of the front panel (including the built-in version) | IP 42 | |
| VACUU·BUS®High power (24V DC) | 4 A | |
| Cable length of external power supply | m | 2 |
| interface | Ethernet / USB (Type A) | |
| Dimensions (length x width x height) | mm | 191 x 127 x 187 |
| weight | kg | 2 |
| NRTL standard | Canada and the United States | |
| ATEX certification | II 3/- G IIC T4 X Internal Atm. only | |
| Product configuration | VACUU · SELECT all-in-one desktop vacuum controller, equipped with solenoid valve, one-way check valve, integrated ceramic diaphragm vacuum sensor and vent valve. Equipped with power supply, can be used at any time, including operation manual. |
VACUU · SELECT - Dimensional Specification Table