Suitable for performance testing of mask bacterial filtration efficiency by metrological verification departments, research institutes, mask production enterprises, and other relevant departments.
YY 0469-2011 Technical Requirements for Medical Surgical Masks Appendix B Bacterial Filtration Efficiency (BFE)
YY/T 0969-2013 Disposable Medical Masks
ASTMF2100, ASTMF2101, EN14683
Q/0212 ZRB003-2011 Bacterial Filtration Efficiency (BFE) Tester for Medical Surgical Masks
Technical Specifications:
1. The use of dual gas path simultaneous comparative sampling method has improved the accuracy of sampling.
2. Negative pressure experimental system to ensure the safety of operators; The negative pressure cabinet is equipped with a peristaltic pump, A and B channels of six stages Anderson; The flow rate of the peristaltic pump can be set;
3. The flow rate of special microbial aerosol generator at night can be set, and the atomization effect is good; Embedded high-speed industrial microcomputer control;
4. 10.4-inch industrial grade high brightness color touch display screen; USB interface, supporting USB data transfer and storage;
5. The cabinet is equipped with high brightness lighting fixtures; Built in leakage protection switch to protect the safety of operators;
6. The inner layer of the cabinet is made of stainless steel and processed as a whole, while the outer layer is made of spray coated cold-rolled sheet. The insulation and flame retardant properties are maintained between the inner and outer layers;
Front switch glass door, convenient for experimental personnel to observe and operate;
7. Detachable bracket with adjustable height; Dual purpose casters for support and mobility.
Main parameters |
parameter range |
resolution |
Maximum allowable error |
A-channel sampling flow |
28.3L/min |
0.1L/min |
Better than ± 2.5% |
B sampling flow rate |
28.3L/min |
0.1L/min |
Better than ± 2.5% |
Spray flow |
(8~10)L/min |
0.1L/min |
Better than ± 5.0% |
Peristaltic pump flow rate |
(0.006~3.0)mL/min |
0.001ml/min |
Better than ± 2.5% |
Pressure in front of A flow meter |
(-20~0)kPa |
0.01kPa |
Better than ± 2.5% |
Pressure in front of B-channel flowmeter |
(-20~0)kPa |
0.01kPa |
Better than ± 2.5% |
Front pressure of spray flowmeter |
(0~300)kPa |
0.1kPa |
Better than ± 2.5% |
Negative pressure in the aerosol chamber |
(-90~-120)Pa |
0.1Pa |
Better than ± 2.0% |
Operating Temperature |
(0~50)℃ |
||
Cabinet negative pressure |
(-50~-200)Pa |
||
Data storage capability |
>100000 sets |
||
Characteristics of high-efficiency air filter |
The filtration efficiency for particles above 0.3um is ≥ 99.99% |
||
Mass median diameter of aerosol generator |
Average particle diameter (3.0 ± 0.3) μ m, geometric standard deviation ≤ 1.5 |
||
Dual channel 6-stage Anderson sampler captures particle size |
Ⅰ级>7 μm,Ⅱ 级 (4.7~7)μm,Ⅲ 级 (3.3~4.7)μm, Ⅳ级 (2.1~3.3)μm,Ⅴ 级 (1.1~2.1)μm, Grade VI (0.6-1.1) μ m |
||
Total number of particles in the positive quality control sampler |
(2200±500)cfu |
||
Ventilation flow rate of negative pressure cabinet |
≥5m3/min |
||
Bracket size |
Height adjustable within 100mm |
||
working power supply |
AC220V±10%,50Hz |
||
instrument noise |
<65dB(A) |
||
Total power consumption |
<1500W |
||