E-mail
13538469076@163.com
-
Phone
13538469076
-
Address
Dongcheng Lixin Jinhui Industrial Park, Dongguan City, Guangdong Province
Guangdong Aisili Testing Instrument Co., Ltd
13538469076@163.com
13538469076
Dongcheng Lixin Jinhui Industrial Park, Dongguan City, Guangdong Province
Produced by Aisili Xenon lamp sun aging test chamber 135384697007 The box body is processed and formed by CNC machine tools, with a beautiful and elegant shape. The box cover is a double-sided flip cover, which is easy to operate.The deviceThe inner liner of the box is made of imported stainless steel plate, and the outer liner of the box is made of A3 steel plate spray painted, which increases the appearance texture and cleanliness.

aging test chamberofProduct Model
Model: XL-1000
Inner box size: 60 × 85 × 80 165 × 195 × 155
Outer box size: 100 × 100 × 100 190 × 210 × 185
aging test chamberProduct Parameters:
Output optical power |
spectral range |
UV output power |
Visible area output power |
50W |
300-2500nm |
2.6W |
19.6W |
50W |
200-2500nm |
6.6W |
17.6W |
power ripple |
Light instability (transient) |
Light instability (period) |
working mode |
± 0.3% (peak to peak) |
Less than ± 0.5% |
Less than ± 1% (10H) |
Electric control/light control mode |
± 0.3% (peak to peak) |
Less than ± 0.5% |
Less than ± 1% (10H) |
Electric control/light control mode |
Light outlet spot diameter |
beam divergence angle |
Filter method |
Lightbox material |
30-63mm |
5 ° -8 ° (depending on the age of the bulb) |
Multi level filtering and turning device |
Non metallic material |
30-63mm |
5 ° -8 ° (depending on the age of the bulb) |
Multi level filtering and turning device |
Non metallic material |
Lightbox volume |
cooling method |
||
120(L)*120(W)*120(H)mm |
Radial heat dissipation module+metal fan |
||
120(L)*120(W)*120(H)mm |
Radial heat dissipation module+metal fan |
||
Working temperature of anode heat sink |
Light bulb lifespan timer |
Lamp lifespan |
|
76 ℃ (Permitted temperature: 112 ℃) |
have |
Not less than 1000H |
|
78 ℃ (Permitted temperature: 112 ℃) |
have |
Not less than 1000H |
aging test chamberElectrical safety
Overheating shutdown protection, fan delay;
Overload and overcurrent protection function;
Overheating shutdown protection, fan delay;

aging test chamberOverload and overcurrent protection function;
The heat dissipation of light sources is influenced by various factors such as wind pressure, air volume, heat dissipation area, and heat dissipation structure; Both axial and radial heat dissipation have their own advantages and disadvantages.
The core parameters of the light source used in photocatalytic research are spectral morphology, photostability, uniformity, and light intensity. The spectral morphology is determined by the characteristics of the light bulb and filter; Light stability is mainly determined by the light bulb and power control system; Uniformity is determined by the optical system; The light intensity is mainly determined by the bulb, optical structure, and power control system.
The collection of optoelectronic control logic, related optoelectronic, circuit components, and optical path systems. Real time light signals of the light bulb are collected by a photodetector with high sensitivity and a certain wavelength response range. After processing with ambient light and stray light, they are converted into electrical signals. Through continuous high-speed collection and comparison, the control signal is transmitted to the circuit output part to adjust the output electrical parameters (gain or compensation) to ensure the stability of the light.
Xenon lamp sun aging test chamber ofpurpose:
Also known as hernia lamp aging test boxSimulate and enhance the aging and damage caused by environmental factors such as solar radiation, temperature, humidity, condensation, and rain in natural climate, observe and analyze the changes in non-metallic materials, organic materials (paint, coatings, rubber, plastics), etc.